Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Frictional force analysis of stent retriever devices using a realistic vascular model: Pilot studyopen access

Authors
Kwak, YoungseokSon, WonsooKim, Byoung-JoonKim, MyungsooYoon, Sang-YoulPark, JaechanLim, JongkyeongKim, JoonwonKang, Dong-Hun
Issue Date
Aug-2022
Publisher
FRONTIERS MEDIA SA
Keywords
acute ischemic stroke; vascular model; frictional retrieval force; Trevo XP; Solitaire 2; Eric 4
Citation
FRONTIERS IN NEUROLOGY, v.13
Journal Title
FRONTIERS IN NEUROLOGY
Volume
13
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/85589
DOI
10.3389/fneur.2022.964354
ISSN
1664-2295
Abstract
Objective: To date, no vascular model to analyze frictional forces between stent retriever devices and vessel walls has been designed to be similar to the real human vasculature. We developed a novel in vitro intracranial cerebrovascular model and analyzed frictional forces of three stent retriever devices. Methods: A vascular mold was created based on digital subtraction angiography of a patient's cerebral vessels. The vascular model was constructed using polydimethylsiloxane (PDMS, Dow Corning, Inc.) as a silicone elastomer. The vascular model was coated on its inner surface with a lubricating layer to create a low coefficient of friction (similar to 0.037) to closely approximate the intima. A pulsatile blood pump was used to produce blood flow inside the model to approximate real vascular conditions. The frictional forces of Trevo XP, Solitaire 2, and Eric 4 were analyzed for initial and maximal friction retrieval forces using this vascular model. The total pulling energy generated during the 3 cm movement was also obtained. Results: Results for initial retrieval force were as follows: Trevo, 0.09 +/- 0.04 N; Solitaire, 0.25 +/- 0.07 N; and Eric, 0.33 +/- 0.21 N. Results for maximal retrieval force were as follows: Trevo, 0.36 +/- 0.07 N; Solitaire, 0.54 +/- 0.06 N; and Eric, 0.80 +/- 0.13 N. Total pulling energy (N.cm) was 0.40 +/- 0.10 in Trevo, 0.65 +/- 0.10 in Solitaire, and 0.87 +/- 0.14 in Eric, respectively. Conclusions: Using a realistic vascular model, different stent retriever devices were shown to have statistically different frictional forces. Future studies using a realistic vascular model are warranted to assess SRT devices.
Files in This Item
There are no files associated with this item.
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lim, Jongkyeong photo

Lim, Jongkyeong
Engineering (기계·스마트·산업공학부(기계공학전공))
Read more

Altmetrics

Total Views & Downloads

BROWSE