Evaluation of the effects of differences in metal artifact type and location on image quality in computed tomography scans

Authors
Kim, SungchulLim, Se-WonChoi, Woo-Keun
Issue Date
Jan-2023
Publisher
Medknow
Keywords
Contrast-to-noise ratio; metal artifact; signal-to-noise ratio; smart metal artifact reduction algorithm; standard algorithm
Citation
Journal of Medical Physics, v.48, no.1, pp 80 - 84
Pages
5
Journal Title
Journal of Medical Physics
Volume
48
Number
1
Start Page
80
End Page
84
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/89083
DOI
10.4103/jmp.jmp_87_22
ISSN
0971-6203
1998-3913
Abstract
Artifacts in computed tomography scans distort anatomical information and prevent an accurate diagnosis. Therefore, this study aims to determine the most effective method for reducing metal-induced artifacts by evaluating the effects of the metal artifact type and location, and the tube voltage on the image quality. Fe and Cu wires were inserted into a Virtual Water (TM) phantom at 6.5 and 11 cm distances from the center point (DPs). The contrast-to-noise ratios (CNRs) and signal-to-noise ratios (SNRs) were calculated to compare the images. The results reveal higher CNRs and SNRs when using standard and Smart metal artifact reduction (Smart MAR) algorithms for Cu and Fe insertions, respectively. The standard algorithm leads to a higher CNR and SNR for Fe and Cu at DPs of 6.5 and 11 cm, respectively. The Smart MAR algorithm provides effective outcomes at voltages of 100 and 120 kVp for wires located at 11 and 6.5 cm DP, respectively. The most effective imaging conditions for MAR is generated by the Smart MAR algorithm with a tube voltage for 100 kVp for Fe located at a DP of 11 cm. MAR can be improved by setting suitable tube voltage conditions according to the type and location of inserted metal.
Files in This Item
Appears in
Collections
보건과학대학 > 방사선학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Sung Chul photo

Kim, Sung Chul
Health Science (Dept.of Radiology)
Read more

Altmetrics

Total Views & Downloads

BROWSE