Material classification using basis material decomposition from spectral X-ray CTopen access

Authors
Jumanazarov, DoniyorAlimova, AsalkhonAbdikarimov, AzamatKoo, JakeoungPoulsen, Henning F.Olsen, Ulrik L.Iovea, Mihai
Issue Date
Nov-2023
Publisher
ELSEVIER
Keywords
Spectral X-ray CT; Data processing methods; Quantitative X-ray characterization; Photon counting detectors; Security screening
Citation
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, v.1056
Journal Title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
Volume
1056
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/89463
DOI
10.1016/j.nima.2023.168637
ISSN
0168-9002
1872-9576
Abstract
Spectral X-ray Computed Tomography (CT) exploits advanced photon counting detectors (PCD) to measure a material's spectrally resolved linear attenuation coefficient (LAC) with the simultaneous spectral acquisition at multiple energy thresholds. We present a method for material classification using spectral CT. The method employs a basis material decomposition model and estimates the effective atomic number (Zeff) from the spectral LAC measurements. Basis material decomposition builds on the fact that the LAC of any material can be well approximated by a linear combination of LACs of basis materials, with known Zeff values at the extremes of the relevant Zeff range. Spectral distortions of the energy spectrum due to the physical interactions between photons and the multi-energy-bin PCD such as charge sharing and photon pileup are corrected by a spectral correction algorithm. The validation of the method has been performed with experimental data acquired with a custom laboratory instrument for spectral CT, examining "real life"phantoms with materials in the range of 6 < Zeff < 15. The classification performance is estimated for different numbers of projections, energy bins and basis materials in LAC decomposition. When using just 12 projections, 15 energy bins and two basis materials, the method gives a relative deviation of 2.2% for Zeff, while this deviation is 5.9% when spectral correction is not used. The classification method is now ready for use in security screening where modern spectral CT systems are employed.
Files in This Item
There are no files associated with this item.
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Koo, JaKeoung photo

Koo, JaKeoung
College of IT Convergence (Department of AI)
Read more

Altmetrics

Total Views & Downloads

BROWSE