Performance evaluation of improved median-modified Wiener filter with segmentation method to improve resolution in computed tomographic images

Authors
Park, JuyoungSong, SeyoungKang, Seong-HyeonLee, Youngjin
Issue Date
Apr-2024
Publisher
KOREAN PHYSICAL SOC
Keywords
Computed tomography; Median modified Wiener filter (MMWF); Image segmentation; Noise reduction; Quantitative evaluation
Citation
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, v.84, no.7, pp 573 - 581
Pages
9
Journal Title
JOURNAL OF THE KOREAN PHYSICAL SOCIETY
Volume
84
Number
7
Start Page
573
End Page
581
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/91665
DOI
10.1007/s40042-024-01020-y
ISSN
0374-4884
1976-8524
Abstract
This study aimed to model and apply an improved median-modified Wiener filter (MMWF) with region growing (RG) segmentation technique in computed tomography (CT) images to improve the noise and blurring effects. To model the improved MMWF, the MATLAB program was used to provide an additional application of median filter and RG techniques from the smoothening process of Wiener filter. In addition, the kernel size of the improved MMWF was set to 7 x 7 and applied to abdominal CT images of the acquired whole body phantom PBU-50 (Kyoto Kagaku, Japan) with tube currents of 50, 100, 200, and 300 mAs, and quantitatively compared to images with conventional MMWF and without MMWF (noisy). Compared with noisy images, the coefficient of variation (COV) and contrast-to-noise ratio (CNR) of the conventional and improved MMWFs improved by approximately 3.41 and 3.32 times, respectively, for the images at all tube current conditions. The improved MMWF showed improved separation between different tissues compared with noisy images even though the smoothening was performed on images with low-dose conditions (50 and 100 mAs). Moreover, a comparative evaluation with other conventional filters (median, Wiener, total variation, and non-local means) demonstrated the improved MMWF performance at low-dose conditions. The COV and CNR of the improved MMWF were overall balanced; however, the improved MMWF performed better than other filters to restore the boundary signal between the two tissues. In conclusion, the improved MMWF with RG technique could acquire CT images with improved characteristics by removing the noise and blurring effects.
Files in This Item
There are no files associated with this item.
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Youngjin photo

Lee, Youngjin
Health Science (Dept.of Radiology)
Read more

Altmetrics

Total Views & Downloads

BROWSE