Detailed Information

Cited 3 time in webofscience Cited 4 time in scopus
Metadata Downloads

Development of a Brain-Computer Interface Toggle Switch with Low False-Positive Rate Using Respiration-Modulated Photoplethysmographyopen access

Authors
Han, Chang-HeeKim, EuijinIm, Chang-Hwan
Issue Date
Jan-2020
Publisher
MDPI
Keywords
electroencephalography; photoplethysmography; asynchronous brain-computer interface; respiration; steady-state visual evoked potential
Citation
SENSORS, v.20, no.2, pp.1 - 13
Indexed
SCIE
SCOPUS
Journal Title
SENSORS
Volume
20
Number
2
Start Page
1
End Page
13
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/11465
DOI
10.3390/s20020348
ISSN
1424-8220
Abstract
Asynchronous brain-computer interfaces (BCIs) based on electroencephalography (EEG) generally suffer from poor performance in terms of classification accuracy and false-positive rate (FPR). Thus, BCI toggle switches based on electrooculogram (EOG) signals were developed to toggle on/off synchronous BCI systems. The conventional BCI toggle switches exhibit fast responses with high accuracy; however, they have a high FPR or cannot be applied to patients with oculomotor impairments. To circumvent these issues, we developed a novel BCI toggle switch that users can employ to toggle on or off synchronous BCIs by holding their breath for a few seconds. Two states-normal breath and breath holding-were classified using a linear discriminant analysis with features extracted from the respiration-modulated photoplethysmography (PPG) signals. A real-time BCI toggle switch was implemented with calibration data trained with only 1-min PPG data. We evaluated the performance of our PPG switch by combining it with a steady-state visual evoked potential-based BCI system that was designed to control four external devices, with regard to the true-positive rate and FPR. The parameters of the PPG switch were optimized through an offline experiment with five subjects, and the performance of the switch system was evaluated in an online experiment with seven subjects. All the participants successfully turned on the BCI by holding their breath for approximately 10 s (100% accuracy), and the switch system exhibited a very low FPR of 0.02 false operations per minute, which is the lowest FPR reported thus far. All participants could successfully control external devices in the synchronous BCI mode. Our results demonstrated that the proposed PPG-based BCI toggle switch can be used to implement practical BCIs.
Files in This Item
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Im, Chang Hwan photo

Im, Chang Hwan
COLLEGE OF ENGINEERING (서울 바이오메디컬공학전공)
Read more

Altmetrics

Total Views & Downloads

BROWSE