Axially slanted laser illumination scheme for direct and accurate Raman spectroscopic determination of gemcitabine concentration in freeze-dried gemcitabine injection powder housed in a glass container

Authors
Lee, YoonjeongKim, JaejinHan, JangheeJeong, HaeseongWoo, Young-AhChung, Hoeil
Issue Date
Aug-2021
Publisher
Elsevier BV
Keywords
Non-sampling; Raman analysis; Gemcitabine injection powder; Axially slanted illumination scheme; Axially perpendicular offset scheme; Glass background suppression
Citation
Analytica Chimica Acta, v.1175, pp 1 - 9
Pages
9
Indexed
SCIE
SCOPUS
Journal Title
Analytica Chimica Acta
Volume
1175
Start Page
1
End Page
9
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/141424
DOI
10.1016/j.aca.2021.338746
ISSN
0003-2670
1873-4324
Abstract
When Raman spectroscopy is employed for a direct in situ determination of ingredient concentration for a product stored in a glass container, minimization of the interfering glass background in the collected spectrum is demanding to secure a more accurate analysis. To meet this request, an axially slanted illumination (ASI) scheme slantingly irradiating laser on the headspace side of a glass container and positioning a detector beneath the container was demonstrated in this study. This ASI scheme was basically designed to increase the distance between the laser illumination spot and detector location to minimize the number of glass photons reaching the detector. The analytical utility of the scheme was evaluated for the determination of gemcitabine concentration (42.9–58.2 wt%) in the gemcitabine injection powder housed in a glass container. Using the ASI scheme, the spectral features of the gemcitabine powder became distinct with only a weak underlying glass background signal. For comparative purpose, when an axially perpendicular offset (APO) scheme perpendicularly irradiating laser on the side wall where the sample was filled was used, the magnitude of glass background was higher, and the most intense gemcitabine peak was largely buried in the glass peak. The accuracy for determination of gemcitabine concentration using the ASI scheme was superior with an error of 0.20 wt%, while 0.33 wt% with employing the APO scheme. Overall, this study demonstrates that the ASI scheme is a potentially versatile Raman spectroscopic tool for fast non-sampling analysis of other products stored in a glass container.
Files in This Item
Go to Link
Appears in
Collections
서울 자연과학대학 > 서울 화학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Chung, Hoeil photo

Chung, Hoeil
COLLEGE OF NATURAL SCIENCES (DEPARTMENT OF CHEMISTRY)
Read more

Altmetrics

Total Views & Downloads

BROWSE