Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

High-throughput residual white blood cell counter enabled by microfluidic cell enrichment and reagent-containing patch integration

Authors
Kim, ByeongyeonShin, SuyeonLee, YujinUm, ChangyongYou, DongwonYun, HoyoungChoi, Sung young
Issue Date
Mar-2019
Publisher
ELSEVIER SCIENCE SA
Keywords
Residual white blood cells; Blood cell counter; Microfluidic cell enrichment; On-chip sample preparation; Hydrogel patch
Citation
SENSORS AND ACTUATORS B-CHEMICAL, v.283, pp.549 - 555
Indexed
SCIE
SCOPUS
Journal Title
SENSORS AND ACTUATORS B-CHEMICAL
Volume
283
Start Page
549
End Page
555
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/148118
DOI
10.1016/j.snb.2018.12.069
ISSN
0925-4005
Abstract
A residual white blood cell (rWBC) count is the remaining concentration of WBCs in leucodepleted blood products, and is an important quality-assurance parameter for blood transfusion. Current cell-counting methods rely on a small test volume that can result in the inherent statistical error involved in counting rWBCs present at low concentrations or manual sample preparation that increases experimental workload. To address these critical challenges, we present a microfluidic rWBC counter that (i) rapidly enriches rWBCs to reduce a sample volume to be examined and thus enable high-throughput cell counting, and (ii) integrates an on-chip cell staining module that contains reagent-containing hydrogel patches and allows the controlled release of a staining reagent. With these novel features, we demonstrated high-throughput rWBC counting by achieving reliable cell counting over a dynamic range of 0.1-26.8 rWBCs/mu L and processing a blood sample at a high throughput of 400 mu L per minute. In addition, we successfully performed sample preparation-free counting of rWBCs by integrating the on-chip staining module into the rWBC counter, offering sample-in-answer-out capability. On the basis of the results, the rWBC counter represents a promising tool for the quality control of WBC depletion and improvement in the safety of blood transfusion.
Files in This Item
Go to Link
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Choi, Sungyoung photo

Choi, Sungyoung
COLLEGE OF ENGINEERING (서울 바이오메디컬공학전공)
Read more

Altmetrics

Total Views & Downloads

BROWSE