Three dimensional neuro-retinal rim thickness and retinal nerve fiber layer thickness using high-definition optical coherence tomography for open-angle glaucoma
- Authors
- Subramaniam, Suresh; Jeoung, Jin Wook; Lee, Won June; Kim, Young Kook; Park, Ki Ho
- Issue Date
- Nov-2018
- Publisher
- SPRINGER JAPAN KK
- Keywords
- End point of Bruch`s membrane; High-definition optical coherence tomography; Three-dimensional neuro-retinal rim thickness mapping; Unilateral open-angle glaucoma; Vitreo-retinal interface
- Citation
- JAPANESE JOURNAL OF OPHTHALMOLOGY, v.62, no.6, pp.634 - 642
- Indexed
- SCIE
SCOPUS
- Journal Title
- JAPANESE JOURNAL OF OPHTHALMOLOGY
- Volume
- 62
- Number
- 6
- Start Page
- 634
- End Page
- 642
- URI
- https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/148943
- DOI
- 10.1007/s10384-018-0620-7
- ISSN
- 0021-5155
- Abstract
- Purpose: To compare the diagnostic capability of three-dimensional (3D) neuro-retinal rim thickness (NRR) with existing optic nerve head and retinal nerve fiber layer (RNFL) scan parameters using high-definition optical coherence tomography (HD-OCT).
Design: Retrospective study.
Methods: Based on the mean deviation (MD) of the Humphrey Field Analyzer (HFA), the 152 subjects were categorized into mild (MD > - 6 dB, 100), moderate (MD - 6 to - 12 dB, 26), and severe (MD < - 12 dB, 26) glaucoma. The HD-OCT values of NRR, RNFL and ganglion cell inner plexiform layer (GCIPL) thicknesses, along with those of other parameters (rim area, disc area) were obtained, and the average NRR thickness was calculated.
Results: For all of the HD-OCT parameters, RNFL thickness showed a higher area under the ROC (AUROC) curve (range: 0.937-1.000) than did NRR thickness (range: 0.827-1.000). There were significant RNFL, NRR, and GCIPL AUROC curve differences among the mild, moderate and severe glaucoma groups. RNFL thickness for mild glaucoma showed a significantly larger area than did NRR thickness [area difference: 0.110 (± 0.025); p value < 0.0001). Furthermore, RNFL relative to NRR thickness yielded higher sensitivity (85-100% vs. 72-100%) and specificity (89-100% vs. 84-100%) for diagnosis of glaucoma.
Conclusion: RNFL thickness remains significantly better than 3D NRR thickness in terms of glaucoma-diagnostic capability in HD-OCT.
- Files in This Item
-
Go to Link
- Appears in
Collections - 서울 의과대학 > 서울 안과학교실 > 1. Journal Articles
![qrcode](https://api.qrserver.com/v1/create-qr-code/?size=55x55&data=https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/148943)
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.