Emergency medical technician-performed point-of-care blood analysis using the capillary blood obtained from skin puncture

Authors
Kim, ChangsunKim, Hansol
Issue Date
Jul-2018
Publisher
W B SAUNDERS CO-ELSEVIER INC
Keywords
Point-of-care systems; Capillary blood; Emergency medical services
Citation
AMERICAN JOURNAL OF EMERGENCY MEDICINE, v.36, no.7, pp.1215 - 1221
Indexed
SCIE
SCOPUS
Journal Title
AMERICAN JOURNAL OF EMERGENCY MEDICINE
Volume
36
Number
7
Start Page
1215
End Page
1221
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/149716
DOI
10.1016/j.ajem.2017.12.025
ISSN
0735-6757
Abstract
Objective Comparing a point-of-care (POC) test using the capillary blood obtained from skin puncture with conventional laboratory tests. Methods In this study, which was conducted at the emergency department of a tertiary care hospital in April–July 2017, 232 patients were enrolled, and three types of blood samples (capillary blood from skin puncture, arterial and venous blood from blood vessel puncture) were simultaneously collected. Each blood sample was analyzed using a POC analyzer (epoc® system, USA), an arterial blood gas analyzer (pHOx®Ultra, Nova biomedical, USA) and venous blood analyzers (AU5800, DxH2401, Beckman Coulter, USA). Twelve parameters were compared between the epoc and reference analyzers, with an equivalence test, Bland-Altman plot analysis and linear regression employed to show the agreement or correlation between the two methods. Results The pH, HCO3, Ca2+, Na+, K+, Cl−, glucose, Hb and Hct measured by the epoc were equivalent to the reference values (95% confidence interval of mean difference within the range of the agreement target) with clinically inconsequential mean differences and narrow limits of agreement. All of them, except pH, had clinically acceptable agreements between the two methods (results within target value ≥ 80%). Of the remaining three parameters (pCO2, pO2 and lactate), the epoc pCO2 and lactate values were highly correlated with the reference device values, whereas pO2 was not. (pCO2: R2 = 0.824, y = − 1.411 + 0.877·x; lactate: R2 = 0.902, y = − 0.544 + 0.966·x; pO2: R2 = 0.037, y = 61.6 + 0.431·x). Conclusion Most parameters, except only pO2, measured by the epoc were equivalent to or correlated with those from the reference method.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 응급의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Chang sun photo

Kim, Chang sun
COLLEGE OF MEDICINE (DEPARTMENT OF EMERGENCY MEDICINE)
Read more

Altmetrics

Total Views & Downloads

BROWSE