Reliable and non-destructive Raman analysis to determine the urea concentration in a cream formulation

Authors
Oh, ChangyongSriphong, LawanKim, MinjungChung, HoeilAsavapichayont, Panida
Issue Date
Mar-2011
Publisher
WILEY
Keywords
urea cream; cream formulation; sample representation; emulsion; wide area illumination
Citation
JOURNAL OF RAMAN SPECTROSCOPY, v.42, no.3, pp.445 - 451
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF RAMAN SPECTROSCOPY
Volume
42
Number
3
Start Page
445
End Page
451
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/151324
DOI
10.1002/jrs.2712
ISSN
0377-0486
Abstract
We present a reliable and nondestructive analytical method for the determination of urea concentration in a pharmaceutical cream formulation using Raman spectroscopy. A pharmaceutical cream is a highly viscous emulsion; therefore, its composition and physical mixing could be inhomogeneous on a microscopic scale. The local environment around the urea could vary, which could influence the molecular vibrations of the urea molecule. As expected, when Raman spectra were collected by focusing the laser onto a tiny area (similar to 2-3 mu m), the position of the urea band at 1003 cm(-1) varied as a result of the microscopic inhomogeneity within the sample. Therefore, acquisition of Raman spectra representative of the entire sample rather than a localized portion of it is very important for the analysis of pharmaceutical creams. Based on the preliminary Raman mapping results of a urea cream, a sample area of at least 750 x 750 mu m should be covered for reliable quantitative analysis. In this study, we used a wide-area illumination scheme capable of covering a sample area of 28.3 mm(2) for Raman spectral collection in order to ensure a reliable representative sample. In addition, to simplify the measurements, Raman spectra of urea creams in plastic bottles were directly collected without further sampling, and partial least squares regression was used for quantitative analysis. The urea concentrations were accurately determined despite the spectral collection being performed through plastic bottles.
Files in This Item
Go to Link
Appears in
Collections
서울 자연과학대학 > 서울 화학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Chung, Hoeil photo

Chung, Hoeil
COLLEGE OF NATURAL SCIENCES (DEPARTMENT OF CHEMISTRY)
Read more

Altmetrics

Total Views & Downloads

BROWSE