Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Primary implant stability in a bone model simulating clinical situations for the posterior maxilla: an in vitro studyopen access

Authors
Han, Ho-ChyulLim, Hyun-ChangHong, Ji-YounAhn, Su-JinHan, Ji-YoungShin, Seung-IlChung, Jong-HyukHerr, YeekShin, Seung-Yun
Issue Date
Aug-2016
Publisher
KOREAN ACAD PERIODONTOLOGY
Keywords
Anatomic models; Dental implants; Maxilla; Maxillary sinus
Citation
JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE, v.46, no.4, pp.254 - 265
Indexed
SCIE
SCOPUS
KCI
Journal Title
JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE
Volume
46
Number
4
Start Page
254
End Page
265
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/154188
DOI
10.5051/jpis.2016.46.4.254
ISSN
2093-2278
Abstract
Purpose: The aim of this study was to determine the influence of anatomical conditions on primary stability in the models simulating posterior maxilla. Methods: Polyurethane blocks were designed to simulate monocortical (M) and bicortical (B) conditions. Each condition had four subgroups measuring 3 mm (M3, B3), 5 mm (M5, B5), 8 mm (M8, B8), and 12 mm (M12, B12) in residual bone height (RBH). After implant placement, the implant stability quotient (ISQ), Periotest value (PTV), insertion torque (IT), and reverse torque (RT) were measured. Two-factor ANOVA (two cortical conditionsxfour RBHs) and additional analyses for simple main effects were performed. Results: A significant interaction between cortical condition and RBH was demonstrated for all methods measuring stability with two-factor ANOVA. In the analyses for simple main effects, ISQ and PTV were statistically higher in the bicortical groups than the corresponding monocortical groups, respectively. In the monocortical group, ISQ and PTV showed a statistically significant rise with increasing RBH. Measurements of IT and RT showed a similar tendency, measuring highest in the M3 group, followed by the M8, the M5, and the M12 groups. In the bicortical group, all variables showed a similar tendency, with different degrees of rise and decline. The B8 group showed the highest values, followed by the B12, the B5, and the B3 groups. The highest coefficient was demonstrated between ISQ and PTV. Conclusions: Primary stability was enhanced by the presence of bicortex and increased RBH, which may be better demonstrated by ISQ and PTV than by IT and RT.
Files in This Item
Appears in
Collections
서울 의과대학 > 서울 치과학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Han, Ji Young photo

Han, Ji Young
COLLEGE OF MEDICINE (DEPARTMENT OF DENTISTRY)
Read more

Altmetrics

Total Views & Downloads

BROWSE