Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Smartwatches as chest compression feedback devices: A feasibility study

Authors
Song, YeongtakChee, YoungjoonOh, JaehoonAhn, ChiwonLim, Tae Ho
Issue Date
Jun-2016
Publisher
ELSEVIER IRELAND LTD
Keywords
CPR; Chest compression; Feedback device; Smartphone; Smartwatch
Citation
RESUSCITATION, v.103, pp.20 - 23
Indexed
SCIE
SCOPUS
Journal Title
RESUSCITATION
Volume
103
Start Page
20
End Page
23
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/154496
DOI
10.1016/j.resuscitation.2016.03.014
ISSN
1748-3107
Abstract
Background: Recently, there have been attempts to use smartphones and smartwatches as the feedback devices to improve the quality of chest compressions. In this study, we compared chest compression depth feedback accuracy between a smartphone and a smartwatch in a hands-only cardiopulmonary resuscitation scenario, using a manikin with a displacement sensor system. Methods: Ten basic life support providers participated in this study. Guided by the chest compression depths displayed on the monitor of a laptop, which received data from the manikin, each participant performed 2min of chest compressions for each target depth (35mm and 55mm) on a manikin while gripping a smartphone and wearing a smartwatch. Participants had a rest of 1h between the instances, and the first target depth was set at random. Each chest compression depth data value from the smartphone and smartwatch and a corresponding reference value from the manikin with the displacement system were recorded. To compare the accuracy between the smartphone and smartwatch, the errors, expressed as the absolute of the differences between the reference and each device, were calculated. Results: At both target depths, the error of the smartwatch were significantly smaller than that of the smartphone (the errors of the smartphone vs. smartwatch at 35mm: 3.4 (1.3) vs. 2.1 (0.8) mm; p=0.008; at 55mm: 5.3 (2.8) vs. 2.3 (0.9) mm; p=0.023). Conclusion: The smartwatch-based chest compression depth feedback was more accurate than smartphone-based feedback.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 응급의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lim, Tae Ho photo

Lim, Tae Ho
COLLEGE OF MEDICINE (DEPARTMENT OF EMERGENCY MEDICINE)
Read more

Altmetrics

Total Views & Downloads

BROWSE