Anti-Reflux Ureteral Stent with Polymeric Flap Valve Using Three-Dimensional Printing: An In Vitro Study

Authors
Park, Chang-JuKim, Hyeon-WooJeong, SangdoSeo, SeungwanPark, YangkyuMoon, Hong SangLee, Jong-Hyun
Issue Date
Aug-2015
Publisher
MARY ANN LIEBERT, INC
Citation
JOURNAL OF ENDOUROLOGY, v.29, no.8, pp.933 - 938
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF ENDOUROLOGY
Volume
29
Number
8
Start Page
933
End Page
938
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/156703
DOI
10.1089/end.2015.0154
ISSN
0892-7790
Abstract
Purpose: This article aims to describe the design of an anti-reflux ureteral stent with a polymeric flap valve and the fabrication methods using three-dimensional (3D) printing. The stent effectively prevents backward flow with a negligible reduction in forward flow. Fabrication of miniaturized valves was easy with high precision and rapid prototyping. Materials and Methods: The proposed stent comprised a 7F Double-J (DJ) stent and a polymeric flap valve. The valve was made of Tango Plus FLX980 and was fabricated using a 3D printer. Two types of stent were prepared for in vitro tests: DJ stents with (1) an uncoated valve (UCV) stent and (2) a parylene C coated valve (PCV) stent for enhanced biocompatibility. The flow characteristics of each stent were evaluated considering flow direction, parylene coating, and stent side holes, and were compared to the intact DJ stent. Results: The forward flow rate for the distal portion of the UCV and PCV stents was 9.8mL/min and 7.8mL/min at applied pressure of 15cm H2O (normal anterograde pressure in patients with stents), respectively. Backward flow rate for the distal portion of the UCV and PCV stents was decreased by 28 times and 8 times at applied pressure of 50cm H2O (maximum bladder pressure), respectively, compared with the distal portion of the intact DJ stent. Forward flow rates of whole stents were 22.2mL/min (UCV stent) and 20.0mL/min (PCV stent) at applied pressure of 15cm H2O, and backward flow rates of whole UCV and PCV stents were decreased by 8.3 times and 4.0 times at applied pressure of 50cm H2O, respectively, compared with the intact DJ stent. Conclusions: The anti-reflux ureteral stent was successfully designed and fabricated using a 3D printer. In vitro studies showed that the stent effectively prevented backward flow while minimizing reduction in forward flow.
Files in This Item
There are no files associated with this item.
Appears in
Collections
서울 의과대학 > 서울 비뇨의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Moon, Hong Sang photo

Moon, Hong Sang
COLLEGE OF MEDICINE (DEPARTMENT OF UROLOGY)
Read more

Altmetrics

Total Views & Downloads

BROWSE