Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Can the new McGrath laryngoscope rival the GlideScope Ranger portable video laryngoscope? A randomized manikin studyopen access

Authors
Kim, WonheeChoi, Hyuk JoongLim, TaehoKang, Bo Seung
Issue Date
Oct-2014
Publisher
W B SAUNDERS CO-ELSEVIER INC
Citation
AMERICAN JOURNAL OF EMERGENCY MEDICINE, v.32, no.10, pp.1225 - 1229
Indexed
SCIE
SCOPUS
Journal Title
AMERICAN JOURNAL OF EMERGENCY MEDICINE
Volume
32
Number
10
Start Page
1225
End Page
1229
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/159039
DOI
10.1016/j.ajem.2014.07.034
ISSN
0735-6757
Abstract
Purpose: We hypothesized that novices would be able to use the McGrath MAC (Aircraft Medical Ltd, Edinburgh, UK) equally as well as the GlideScope Ranger (Verathon, Inc, Bothell, WA) for intubation in regular simulated airways. Methods: We performed a prospective, randomized crossover study of 39 medical students using the McGrath MAC, GlideScope Ranger, and Macintosh in a manikin with 2 normal airways. The primary outcome was the intubation time. Secondary outcomes included the success rates and the overall glottic view of the 3 laryngoscopes. Results: The mean intubation times for each attempt with the McGrath MAC were 30.8 +/- 16.9 seconds or less and did not differ significantly from those obtained with the GlideScope Ranger or the Macintosh in both airway scenarios (P = .18; P = .49). The mean success rates at each attempt with the McGrath MAC were 82.0% +/- 38.8% or more, equal to the Macintosh and the GlideScope Ranger in both scenarios (P = .026; P = .72) except during the first intubation attempt in a normal airway (P =. 008). The median grades of the glottic view visible at each intubation attempt with the McGrath Mac were Cormack-Lehane grade 1 (scenario 1: interquartile range, 1-1; scenario 2: interquartile range, 1-2), which was significantly better than the Macintosh laryngoscope in both scenarios. However, the McGrath Mac did not produce a better glottic view than the GlideScope Ranger with either scenario. Conclusions: The intubation performance of novices using the McGrath MAC was equal to their performance using the GlideScope Ranger in regular simulated airways.
Files in This Item
Appears in
Collections
서울 의과대학 > 서울 응급의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lim, Tae Ho photo

Lim, Tae Ho
COLLEGE OF MEDICINE (DEPARTMENT OF EMERGENCY MEDICINE)
Read more

Altmetrics

Total Views & Downloads

BROWSE