Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Comparison of Sterilization of Reusable Endoscopic Biopsy Forceps by Autoclaving and Ethylene Oxide Gas

Authors
Yoon, Jai HoonYoon, Byung ChulLee, Hang LakLee, Jun KyuKim, Yong-TaeLee, Dong HoChoi, Il JuLee, Don HaengKim, Dong Hee
Issue Date
Feb-2012
Publisher
SPRINGER
Keywords
Biopsy forceps; Sterilization; Equipment reuse; Ethylene oxide; Autoclave
Citation
DIGESTIVE DISEASES AND SCIENCES, v.57, no.2, pp.405 - 412
Indexed
SCIE
SCOPUS
Journal Title
DIGESTIVE DISEASES AND SCIENCES
Volume
57
Number
2
Start Page
405
End Page
412
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/166344
DOI
10.1007/s10620-011-1884-7
ISSN
0163-2116
Abstract
Every country has standardized reprocessing guidelines for reducing the risk of microorganism transmission via reusable biopsy forceps. Sterilization is performed either by autoclaving or with the use of ethylene oxide (EO) gas. However, there are no clear standard global recommendations. The aim of this study was to determine whether EO gas or autoclaving is a safer and more effective method for the sterilization of reusable forceps. This was a prospective study conducted at multiple tertiary referral centers. Seventy reusable biopsy forceps that had been reused at least 20 times each were collected from six endoscopy centers. In all, 61 forceps from five centers were sterilized using EO gas, and the nine forceps from the remaining center were placed in an autoclave. We performed real-time polymerase chain reaction (RT-PCR) for Mycobacterium tuberculosis and hepatitis B virus and performed bacterial cultures on the reusable forceps, which were cut into 2- to 3-cm sections. The forceps were also scanned with an electron microscope (EM) to detect surface damage and contamination. Escherichia coli bacteria were cultured from 2 of the 61 (3.3%) reusable biopsy forceps sterilized with EO gas. On EM scanning, abundant debris and tissue materials remained on the cup surfaces of the reused biopsy forceps and on their inner wires. No microorganisms were found on the autoclaved forceps. Sterilization with EO gas may be inadequate because the complicated structure of the forceps may interfere with sterilization. Therefore, for optimum safety, reusable biopsy forceps should be sterilized by autoclaving.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 내과학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Hang Lak photo

Lee, Hang Lak
COLLEGE OF MEDICINE (DEPARTMENT OF INTERNAL MEDICINE)
Read more

Altmetrics

Total Views & Downloads

BROWSE