Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Incidence and Factors Associated with Squeaking in Alumina-on-Alumina THAopen access

Authors
Choi, Il-YongKim, Yee-SukHwang, Kyu-TaeKim, Young-Ho
Issue Date
Dec-2010
Publisher
LIPPINCOTT WILLIAMS & WILKINS
Citation
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, v.468, no.12, pp.3234 - 3239
Indexed
SCIE
SCOPUS
Journal Title
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH
Volume
468
Number
12
Start Page
3234
End Page
3239
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/173388
DOI
10.1007/s11999-010-1394-5
ISSN
0009-921X
Abstract
Background The alumina-on-alumina bearing surface, which has a high wear resistance and a good biocompatibility, is widely used in THA but recently has been associated with squeaking. While various authors have reported factors associated with squeaking, they remain poorly understood. Questions/purposes To contribute to the debate on squeaking we therefore asked the following questions: (1) What is the incidence of squeaking in alumina-on-alumina THA? (2) What factors are associated with squeaking in alumina bearings in our practice? Methods We retrospectively reviewed 168 patients (173 hips) who had primary alumina-on-alumina THAs. The mean age of the patients was 53 years (range, 18 to 81 years). Minimum followup was 5.6 years (average, 7.3 years; range, 5.6-9.4 years). All patients were evaluated clinically and radiographically with attention to periprosthetic osteolysis, squeaking, and ceramic fracture. When the patient reported squeaking, we determined the onset, reproducibility, and activities associated with the squeaking. We recorded patient (gender, age, height, weight, and body mass index) and surgical factors (abduction angle of cup, size and length of ceramic head component, and diameter of cup in the implant). Results Eight of the 168 patients (5%) had squeaking hips. Squeaking was more common in males and in those with large ceramic heads. There were no complications or revisions in the squeaking group. One ceramic liner fracture was associated with trochanteric nonunion. Conclusions When recommending alumina-on-alumina bearing surfaces to patients they should be clearly informed of the possibility of squeaking. Patients with risk factors for squeaking should be followed at regular intervals.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 정형외과학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Yee suk photo

Kim, Yee suk
COLLEGE OF MEDICINE (DEPARTMENT OF ORTHOPEDIC SURGERY)
Read more

Altmetrics

Total Views & Downloads

BROWSE