Abstract
본 연구는 GC/FID에서 염화유기용제(PCE와 TCE)와 그 부산물(cis-DCE, VC, 그리고 Ethylene)의 분석을 위하여 액-액추출법의 대안인 SPME방법을 연구하였다. 실험인자로써 fiber의 종류, 흡착 및 탈착시간, headspace의 부피, 염석효과, 그리고 교반효과에 대하여 실험을 실시하였다. 염화유기용제와 부산물의 분석을 위하여 fiber는 CAR/PDMS, 흡착과 탈착시간은 250℃에서 20분과5분, 염첨가량 25%와 교반이 병행되었을 때에 최적의 분석조건으로 도출되었다. CAR/PDMS fiber에서 분석물의 추출감도는 PCE>TCE>cis-DCE>VC>Ethylene순으로 조사되었다. 염화유기용제와 부산물의 농도 10μg/L에서 500μg/L까지의 직선상의 상관계수(R2)는 0.912에서 0.999로 조사되었으며, 500μg/L(5회 분석)일 때에 상대표준편차(% RSD)는 2.1%에서 3.6%로 나타났다. 마지막으로본 연구에서 염화유기용제와 부산물의 검출한계(LOD)는 0.5μg/L에서 10μg/L로 관측되었다.
In this study, Solid-phase microextraction (SPME) with GC/FID was studied as a possible alternative to liquid-liquid extraction for the analysis of chlorinated solvents (PCE and TCE) and these by-products (cis-DCE, VC, and Ethylene). Experimental parameters affecting the SPME process (such as kind of fibers, adsorption time, desorption time, volume ratio of sample to headspace,salt addition, and magnetic stirring) were optimized. Experimental parameters such as CAR/PDMS, adsorption time of 20 min,desorption time of 5 min at 250℃, headspace volume of 50mL, sodium chloride (NaCl) concentration of 25% combined with magnetic stirring were selected in optimal experimental conditions for analysis of chlorinated solvents and these by-products. The general affinity of analytes to CAR/PDMS fiber was high in the order PCE>TCE>cis-DCE>VC>Ethylene. The linearity of R2 for chlorinated solvents and these by-products was from 0.912 to 0.999 when analyte concentrations range from 10μg/L to 500μg/L, respectively. The relative standard deviation (% RSD) were from 2.1% to 3.6% for concentration of 500μg/L (n=5), respectively. Finally, the limited of detection (LOD) observed in our study for chlorinated solvents and these by-products were from 0.5μg/L to 10μg/L,respectively.
Files in This Item
Go to Link
Appears in
Collections
서울 공과대학 > 서울 건설환경공학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Park, Jae Woo photo

Park, Jae Woo
COLLEGE OF ENGINEERING (DEPARTMENT OF CIVIL AND ENVIRONMENTAL ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE