Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

물흡수선량 표준에 기반한 192Ir 근접치료 선원 교정 시 원통형 이온함의 이온함 간 변화Chamber to Chamber Variations of a Cylindrical Ionization Chamber for the Calibration of an 192Ir Brachytherapy Source Based on an Absorbed Dose to Water Standards

Other Titles
Chamber to Chamber Variations of a Cylindrical Ionization Chamber for the Calibration of an 192Ir Brachytherapy Source Based on an Absorbed Dose to Water Standards
Authors
Kim, Seong HoonHuh, HyundoChoi, Sang HyunKim, Chan HyeongMin, Chul HeeShin, Dong OhChoi, Jinho
Issue Date
Mar-2009
Publisher
한국의학물리학회
Keywords
Beam quality correction factor; Absorbed dose to water; Ionization chamber; Primary standards; 빔선질 교정인자; 물흡수선량; 이온함; 일차 표준
Citation
의학물리, v.20, no.1, pp.7 - 13
Indexed
KCI
Journal Title
의학물리
Volume
20
Number
1
Start Page
7
End Page
13
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/177090
ISSN
2508-4445
Abstract
This work is for the preliminary study for the calibration of an 192Ir brachytherapy source based on an absorbed dose to water standards. In order to calibrate brachytherapy sources based on absorbed dose to water standards using a clyndirical ionization chamber, the beam quality correction factor kQ,Q0 is needed. In this study kQ,Q0s were determined by both Monte carlo simulation and semiexperimental methods because of the realistic difficulties to use primary standards to measure an absolute dose at a specified distance. The 5 different serial numbers of the PTW30013 chamber type were selected for this study. While chamber to chamber variations ran up to maximum 4.0% with the generic kgenQ,Q0, the chamber to chamber variations were within a maximum deviation of 0.5% with the individual kindQ,Q0. The results show why and how important ionization chambers must be calibrated individually for the calibration of 192Ir brachytherapy sources based on absorbed dose to water standards. We hope that in the near future users will be able to calibrate the brachytherapy sources in terms of an absorbed dose to water, the quantity of interest in the treatment, instead of an air kerma strength just as the calibration in the high energy photon and electron beam.
Files in This Item
Go to Link
Appears in
Collections
서울 공과대학 > 서울 원자력공학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Chan Hyeong photo

Kim, Chan Hyeong
COLLEGE OF ENGINEERING (DEPARTMENT OF NUCLEAR ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE