Determination of Optimal Energy Window for Measurement of Prompt Gammas from Proton Beam by Monte Carlo Simulations
- Authors
- Min, Chul Hee; Park, Jang Guen; An, So Hyun; Kim, Chan Hyeong
- Issue Date
- Jun-2008
- Publisher
- Atomic Energy Society of Japan/Nihon Genshiroku Gakkai
- Keywords
- proton therapy; proton range; prompt gammas; energy window
- Citation
- Journal of Nuclear Science and Technology, pp 28 - 31
- Pages
- 4
- Indexed
- SCIE
SCOPUS
- Journal Title
- Journal of Nuclear Science and Technology
- Start Page
- 28
- End Page
- 31
- URI
- https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/178558
- DOI
- 10.1080/00223131.2008.10875777
- ISSN
- 0022-3131
1881-1248
- Abstract
- The range of a proton beam in a patient can be determined by scanning the distribution of the prompt gammas emitted from the beam passage. However, this method suffers from a high level of background gammas, especially in the case of high energy proton beams. The present study determined the optimal energy window for selective measurement of the prompt gammas, effectively discriminating background gammas, for a prompt gamma scanning system. To that end, the energy spectra of the prompt and background gammas were calculated by transporting the protons and other secondary particles with MCNPX. A detailed analysis of these spectra revealed that the optimal energy window is 4-10 MeV. The application of the energy window to simulated and measured data confirmed that the range of a proton beam in a patient can be determined much more accurately by employing the optimal energy window.
- Files in This Item
-
Go to Link
- Appears in
Collections - 서울 공과대학 > 서울 원자력공학과 > 1. Journal Articles

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.