A Baseline-Tracking Single-Channel I/Q Impedance Plethysmogram IC for Neckband-Based Blood Pressure and Cardiovascular Monitoring

Authors
Park, Chan SamKim, HyunjoongLee, KwangmukKeum, Dae SikJang, Dong PyoKim, Jae Joon
Issue Date
Jul-2023
Publisher
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
Keywords
Baseline tracking; blood pressure (BP); impedance plethysmogram (IPG); neckband; phase synchronizer; pulse arrival time (PAT)
Citation
IEEE JOURNAL OF SOLID-STATE CIRCUITS, v.58, no.7, pp.2040 - 2052
Indexed
SCIE
SCOPUS
Journal Title
IEEE JOURNAL OF SOLID-STATE CIRCUITS
Volume
58
Number
7
Start Page
2040
End Page
2052
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/191094
DOI
10.1109/JSSC.2022.3221478
ISSN
0018-9200
Abstract
An impedance plethysmogram (IPG) IC with wide-range fine baseline tracking capability is proposed to provide continuous blood pressure (BP) and cardiovascular monitoring just by wearing a neckband wearable device. For the baseline tracking, a mixed-mode baseline cancellation (MM-BC) scheme is proposed to achieve wide dynamic range (DR) and high SNR performance, where an artifact-detecting continuous-time (CT)-Delta Sigma ADC is included together for high-resolution conversion. For cost-effective IPG design, a phase synchronizer concept is introduced to provide I/Q detection capability with single readout channel, supporting both configurations of two-electrode and four-electrode. A proposed readout integrated circuit (ROIC) prototype for the proposed IPG was fabricated, where an electrocardiogram (ECG) readout path is integrated together. The proposed IPG schemes were experimentally verified to achieve good performances of 145.2-dB DR and 103.5-dB SNR. For system-level feasibility, a neckband device prototype was also manufactured, and its cardiovascular monitoring functionality was functionally verified. Based on its IPG and ECG measurements, continuous BP measurement based on the pulse arrival time (PAT) was experimentally verified with a reference BP device.
Files in This Item
Go to Link
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Jang, Dong Pyo photo

Jang, Dong Pyo
GRADUATE SCHOOL OF BIOMEDICAL SCIENCE AND ENGINEERING (서울 생체의공학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE