Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Comparisons of an automated oscillometric device with a hybrid manual auscultatory device for the Korea National Health and Nutrition Examination Surveyopen access

Authors
Kim, Yu-MiPark, Sang HyeonShin, JinhoSung, Ki-ChulIhm, Sang-HyunKim, Seong HeonKim, Dae-HeePark, Sang MinCho, In JeongOh, Kyung WonLee, Eun Mi
Issue Date
May-2024
Publisher
Wiley-Blackwell
Keywords
agreement; blood pressure determination; health survey; hypertension; sphygmomanometers
Citation
Journal of Clinical Hypertension, v.26, no.5, pp 532 - 542
Pages
11
Indexed
SCIE
SCOPUS
Journal Title
Journal of Clinical Hypertension
Volume
26
Number
5
Start Page
532
End Page
542
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/202135
DOI
10.1111/jch.14805
ISSN
1524-6175
1751-7176
Abstract
This study evaluated an oscillometric device (OD), Microlife WatchBP Office AFIB, and a hybrid manual auscultatory device (AD), Greenlight 300TM, to determine a suitable blood pressure (BP) measurement device for the Korea National Health and Nutrition Examination Survey in a mercury-free context. Adhering to the 2018 Universal Standard's suggested consensus, the study involved 800 subjects (mean age 51.2 +/- 17.5 years; 44.3% male), who underwent triplicate BP measurements following 5 min of rest in a randomized order (OD-first: 398 participants; AD-first: 402 participants). BP difference was calculated as OD value minus AD value, with results stratified by measurement sequence. The overall BP difference and tolerable error probability were -1.1 +/- 6.5/-2.6 +/- 4.9 mmHg and 89.2%/92.5% for systolic/diastolic BP (SBP/DBP), respectively. Lin's concordance correlation coefficient was 0.907/0.844 for SBP/DBP (OD-first/AD-first: 0.925/0.892 for SBP, 0.842/0.845 for DBP). The overall agreement for hypertension (BP >= 140 and/or 90 mmHg) was 0.71 (p < 0.0001), and the OD underestimated the overall hypertension prevalence by 5.1%. Analysis of the AD-first data revealed a lower level of agreement compared to the OD-first data; however, the observed blood pressure difference adhered to Criterion 1 of the 2018 Universal Standard. Microlife met the Criterion 1 of 2018 Universal Standard but underestimated the prevalence of hypertension. The BP discrepancy increased with higher BP levels, male sex, and smaller AC. With increasing age, the discrepancy decreased for SBP and increased for DBP.
This study evaluated an oscillometric device (OD), Microlife WatchBP Office AFIB, and a hybrid manual auscultatory device (AD), Greenlight 300TM, to determine a suitable blood pressure (BP) measurement device for the Korea National Health and Nutrition Examination Survey in a mercury-free context. Adhering to the 2018 Universal Standard’s suggested consensus, the study involved 800 subjects (mean age 51.2 ± 17.5 years; 44.3% male), who underwent triplicate BP measurements following 5 min of rest in a randomized order (OD-first: 398 participants; AD-first: 402 participants). BP difference was calculated as OD value minus AD value, with results stratified by measurement sequence. The overall BP difference and tolerable error probability were -1.1 ± 6.5/-2.6 ± 4.9 mmHg and 89.2%/92.5% for systolic/diastolic BP (SBP/DBP), respectively. Lin’s concordance correlation coefficient was 0.907/0.844 for SBP/DBP (OD-first/AD-first: 0.925/0.892 for SBP, 0.842/0.845 for DBP). The overall agreement for hypertension (BP ≥ 140 and/or 90 mmHg) was 0.71 (p < 0.0001), and the OD underestimated the overall hypertension prevalence by 5.1%. Analysis of the AD-first data revealed a lower level of agreement compared to the OD-first data; however, the observed blood pressure difference adhered to Criterion 1 of the 2018 Universal Standard. Microlife met the Criterion 1 of 2018 Universal Standard but underestimated the prevalence of hypertension. The BP discrepancy increased with higher BP levels, male sex, and smaller AC. With increasing age, the discrepancy decreased for SBP and increased for DBP.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 내과학교실 > 1. Journal Articles
서울 의과대학 > 서울 예방의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Shin, Jinho photo

Shin, Jinho
서울 의과대학 (DEPARTMENT OF INTERNAL MEDICINE)
Read more

Altmetrics

Total Views & Downloads

BROWSE