DESIGN OF A SIDE-SHIELDED WIRELESS POWER TRANSFER COIL TO ACHIEVE INSENSITIVITY AGAINST THE HUMAN BODY EFFECT

Authors
Lee, JuneseokSong, Keum-SuChoi, Jaehoon
Issue Date
May-2016
Publisher
WILEY
Keywords
wireless power transfer; implantable medical device; side-shielded coil; human body effect
Citation
MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, v.58, no.5, pp.1169 - 1173
Indexed
SCIE
SCOPUS
Journal Title
MICROWAVE AND OPTICAL TECHNOLOGY LETTERS
Volume
58
Number
5
Start Page
1169
End Page
1173
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/23094
DOI
10.1002/mop.29735
ISSN
0895-2477
Abstract
We propose a novel side-shielded wireless power transfer coil for implantable medical devices. The side-shielded coil consists of a resonator and a side-shield structure, which is partially wrapped around the resonator. The side-shield structure isolates the proposed coil from the human body effect and efficiently transfers power to the receiving coil. In addition, the proposed coil has a specific absorption rate approximately four times lower than that of a conventional spiral wireless power transfer coil.