Comparison of the accuracy and errors of blood pressure measured by 2 types of non-mercury sphygmomanometers in an epidemiological surveyopen access

Authors
Choi, SeongllKim, Yu-MiShin, JinhoLim, Young-HyoChoi, Sung-YongChoi, Bo-YoulOh, Kyung-WonLee, Hyung-MinWoo, Kyung-Ji
Issue Date
Jun-2018
Publisher
LIPPINCOTT WILLIAMS & WILKINS
Keywords
auscultation; blood pressure; diagnostic errors; electronic; epidemiological factors; sphygmomanometers
Citation
MEDICINE, v.97, no.25
Indexed
SCIE
SCOPUS
Journal Title
MEDICINE
Volume
97
Number
25
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/2366
DOI
10.1097/MD.0000000000010851
ISSN
0025-7974
Abstract
A few studies have compared auscultation and oscillometric devices with the mercury sphygmomanometer (MS) reference values for blood pressure (BP) measurement in an epidemiologic survey. Four trained observers recorded BP measurements from 766 subjects from general Korean population in 2014 and 2015. Measurements were repeated 3 times for each device, alternately using an MS and 2 electronic devices (EDs; Greenlight 300 vs Omron HEM-907), together with a randomized device sequence. The BP measurement difference was defined as BP measured by MS minus BP obtained by ED, and the absolute error as the absolute value of the difference. Mean differences in systolic BP (SBP) were -0.52 and -0.62 mmHg and those of diastolic BP (DBP) were -0.78 and 6.23 mmHg (P < .01) in the Greenlight and Omron device group, respectively. The concordance correlation coefficients were 0.97 and 0.94 for SBP and 0.95 and 0.76 for DBP in the Greenlight and Omron group, respectively (P < .05). Kappa values for the Joint National Committee 7 BP classification were 0.84 and 0.74 for Greenlight and Omron group, respectively. The prevalence of normotension, prehypertension and hypertension were 53.5%, 33.9%, and 12.5% with the MS and 59.8%, 29.0%, and 11.2% with the ED in the Omron group (P = .03, McNemar test), whereas they were insignificant in the Greenlight group. The Greenlight 300 may be a good alternative to the MS, and the Omron HEM-907 has good accuracy in SBP measurement. Due to the measurement error in DBP, Omron HEM-907 was inferior to the Greenlight device.
Files in This Item
Appears in
Collections
서울 의과대학 > 서울 내과학교실 > 1. Journal Articles
서울 의과대학 > 서울 예방의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lim, Young-Hyo photo

Lim, Young-Hyo
COLLEGE OF MEDICINE (DEPARTMENT OF INTERNAL MEDICINE)
Read more

Altmetrics

Total Views & Downloads

BROWSE