Detailed Information

Cited 1 time in webofscience Cited 2 time in scopus
Metadata Downloads

Feasibility of discrimination of gall bladder (GB) stone and GB polyp using voltage-applied SERS measurement of bile juice samples in conjunction with two-trace two-dimensional (2T2D) correlation analysis

Authors
Vu, Tung DuySohng, WoosukJang, EunjinChoi, DonghoChung, Hoeil
Issue Date
Feb-2021
Publisher
ROYAL SOC CHEMISTRY
Citation
ANALYST, v.146, no.3, pp.1091 - 1098
Indexed
SCIE
SCOPUS
Journal Title
ANALYST
Volume
146
Number
3
Start Page
1091
End Page
1098
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/8016
DOI
10.1039/d0an02115f
ISSN
0003-2654
Abstract
Voltage-applied SERS measurement of bile juice in conjunction with two-trace two-dimensional (2T2D) correlation analysis was demonstrated as a potential tool to enhance discrimination of gall bladder (GB) stone and GB polyp. When SERS spectra of the aqueous phases extracted from raw bile juice samples were measured with applying external voltage from -300 to +300 mV (100 mV intervals), subsequent spectral variations of the adsorbed components (bilirubin-containing compounds) on the SERS substrate were minute, and discrimination of the two GB diseases in a principal component score domain was difficult. Therefore, 2T2D correlation analysis, effectively identifying asynchronous (dissimilar) spectral behaviors in the voltage-induced SERS spectra, was used to improve the discrimination. When two spectra of a sample collected with application of +100 and +300 mV were adopted, the features of subsequent 2T2D slice spectra were characteristic, and discrimination of GB stone and GB polyp substantially improved. External voltage application and recognition of the voltage-induced spectral features by 2T2D correlation analysis were key factors for the improvement. Since the demonstrated method relied on only a few SERS-active compounds, infrared (IR) spectroscopy featuring all the present components in the samples was also evaluated for comparison. However, the IR-based discrimination was inferior because the metabolite compositions in the samples between the GB diseases were not noticeably different.
Files in This Item
Go to Link
Appears in
Collections
서울 자연과학대학 > 서울 화학과 > 1. Journal Articles
서울 의과대학 > 서울 외과학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Chung, Hoeil photo

Chung, Hoeil
COLLEGE OF NATURAL SCIENCES (DEPARTMENT OF CHEMISTRY)
Read more

Altmetrics

Total Views & Downloads

BROWSE