Wavelength discrimination (WLD) TOF-PET detector with DOI information

Authors
Ullah, Muhammad NasirPratiwi, EvaPark, Jin HoLee, KisungChoi, HojongYeom, Jung-Yeol
Issue Date
Mar-2020
Publisher
IOP PUBLISHING LTD
Keywords
phoswich detector; optical filter; emission wavelength; depth of interaction; time-of-flight positron emission tomography
Citation
PHYSICS IN MEDICINE AND BIOLOGY, v.65, no.5
Journal Title
PHYSICS IN MEDICINE AND BIOLOGY
Volume
65
Number
5
URI
https://scholarworks.bwise.kr/kumoh/handle/2020.sw.kumoh/25968
DOI
10.1088/1361-6560/ab6579
ISSN
0031-9155
1361-6560
Abstract
Depth-of-interaction (DOI) encoding can contribute to improving spatial resolution uniformity and sensitivity in positron-emission-tomography (PET) scanners. In addition, time-of-flight (TOF) PET scanners with DOI encoding have received considerable interest because of their potential for improving the spatial resolution, sensitivity, and image quality of the overall system. In this study, a new DOI detector configuration utilizing scintillators' emission wavelength is proposed, and experimental results on the energy, timing, and DOI performance of the detector are provided. The DOI information from the proposed phoswich-type detector can be acquired at the detector level without complex signal processing by utilizing a single optical filter with customized optical properties. For this, we used either a short pass filter (SPF) or a long pass filter (LPF) that allows light photons of a specific wavelength to pass. The two-layered phoswich detector was configured with two scintillators with different photon-emission spectra. In this study, we used Ce:GAGG (3 mm x 3 mm x 10 mm) and LYSO:Ce (3 mm x 3 mm x 10 mm) as the top and bottom layer scintillators, respectively. A digital silicon photomultiplier (dSiPM) was used as the photosensor and for data acquisition. The phoswich detector was placed in the center of two dSiPM pixels, where one of the dSiPM pixels was covered with the optical filter, and the light guide was placed on the other pixel. The detector was tested for energy, timing, and DOI encoding performance. When an SPF was used, the energy resolutions of 16.2% and 11.8% were achieved for the Ce:GAGG (top layer) and LYSO:Ce (bottom layer) respectively without correcting for saturation effect. With a small (3 mm x 3 mm x 5 mm) LYSO crystal as the reference detector, CRTs (coincidence-resolving times) of 338 ps and 244 ps were recorded for the top and bottom layers respectively. The detector configuration also provides an excellent DOI-separation figure-of-merit (FoM) value of 1.9. In the case of LPF, the energy resolutions of 12.0% and 12.9% were achieved for the Ce:GAGG (top layer) and LYSO:Ce (bottom layer), respectively. CRTs (coincidence resolving times) of 314 ps and 263 ps were recorded for the top and bottom layers, respectively. The DOI-separation FoM value of 1.5 was achieved in this setup. Results show that the proposed method can provide excellent discrete DOI positioning accuracy without compromising the timing performance of the detector.
Files in This Item
There are no files associated with this item.
Appears in
Collections
Department of Medical IT Convergence Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE