Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

A Comparative Quantitative Analysis of IDEAL (Iterative Decomposition of Water and Fat with Echo Asymmetry and Least Squares Estimation) and CHESS (Chemical Shift Selection Suppression) Technique in 3.0T Musculoskeletal MRI

Authors
Kim, Myoung-HoonCho, Jae-HwanShin, Seong-GyuDong, Kyung-RaeChung, Woon-KwanPark, Tae-HyunAhn, Jae-OukPark, Cheol-SooJang, Hyon-CholKim, Yoon-Shin
Issue Date
Jun-2012
Publisher
한국자기학회
Keywords
CHESS; IDEAL; susceptibility artifacts; homogeneous fat suppression
Citation
Journal of Magnetics, v.17, no.2, pp 145 - 152
Pages
8
Journal Title
Journal of Magnetics
Volume
17
Number
2
Start Page
145
End Page
152
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/15121
DOI
10.4283/JMAG.2012.17.2.145
ISSN
1226-1750
2233-6656
Abstract
Patients who underwent hip arthroplasty using the conventional fat suppression technique (CHESS) and a new technique (IDEAL) were compared quantitatively to assess the effectiveness and usefulness of the IDEAL technique. In 20 patients who underwent hip arthroplasty from March 2009 to December 2010, fat suppression T2 and T1 weighted images were obtained on a 3.0T MR scanner using the CHESS and IDEAL techniques. The level of distortion in the area of interest, the level of the development of susceptibility artifacts, and homogeneous fat suppression were analyzed from the acquired images. Quantitative analysis revealed the IDEAL technique to produce a lower level of image distortion caused by the development of susceptibility artifacts due to metal on the acquired images compared to the CHESS technique. Qualitative analysis of the anterior area revealed the IDEAL technique to generate fewer susceptibility artifacts than the CHESS technique but with homogeneous fat suppression. In the middle area, the IDEAL technique generated fewer susceptibility artifacts than the CHESS technique but with homogeneous fat suppression. In the posterior area, the IDEAL technique generated fewer susceptibility artifacts than the CHESS technique. Fat suppression was not statistically different, and the two techniques achieved homogeneous fat suppression. In conclusion, the IDEAL technique generated fewer susceptibility artifacts caused by metals and less image distortion than the CHESS technique. In addition, homogeneous fat suppression was feasible. In conclusion, the IDEAL technique generates high quality images, and can provide good information for diagnosis.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medical Sciences > Department of Medical IT Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Ahn, Jae Ouk photo

Ahn, Jae Ouk
College of Software Convergence (의료IT공학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE