휴대용 광이온화 검출기의 정확도, 정밀도 및 기기간 차이에 대한 실험실 평가Laboratory Evaluation of the Accuracy, Precision, and Inter-instrumental Variance of a Portable Photoionization Detector

Other Titles
Laboratory Evaluation of the Accuracy, Precision, and Inter-instrumental Variance of a Portable Photoionization Detector
Authors
최동민최영은윤충식이광원이윤근이익모박정임
Issue Date
2012
Publisher
한국산업보건학회
Keywords
Portable Photoionization Detector; Evaluation; Accuracy; Calibration
Citation
한국산업보건학회지, v.22, no.3, pp.200 - 208
Journal Title
한국산업보건학회지
Volume
22
Number
3
Start Page
200
End Page
208
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/15621
ISSN
2384-132x
Abstract
Objectives: This study investigated the performance of three separate units of a portable photoionization detector (PID,ppb-RAE 3000) for measuring volatile organic compounds (VOCs) in a laboratory. Methods: A laboratory evaluation of the accuracy, precision, and inter-instrumental variance of three separate units of a portable PID (ppb-RAE 3000) was performed. The evaluation was based on the preparation of a test air sample of known toluene or ethylacetate concentration in a Tedlar® bag. The test air sample was monitored and data were logged consecutively by the three PIDs. A certified gas of 50 ppm toluene was also monitored during the test to ensure the reliability of the generated test air sample. Four different concentrations ranging from 0.1 to 2 TLV were used and a series of five measurements for each concentration level was performed. The accuracy was evaluated using National Institute for Occupational Safety and Health (NIOSH) criteria. results: The results from the oldest ppb-RAE3000 unit among the three test units generally fell outside the NIOSH recommended accuracy criteria of ±25%, whereas the other two units produced results which were acceptable at, or greater than, 25 ppm of toluene, or 0.5 TLV. These units also met the NIOSH criteria for some ethylacetate measurements but the results were not consistent. conclusions: Considering the inconsistent performance of these ppb-RAE 3000 units, this device may not be appropriate for use as an alternative to the standard measurement methods. However, it can serve good survey instruments to identify exposure sources or concentration profiles. For all applications, the ppb-RAE 3000 should be used with frequent calibration checks, additional validation using a reference material, and careful maintenance.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Natural Sciences > Department of Environmental Health Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Park, Jeong im photo

Park, Jeong im
College of Natural Sciences (Department of Environmental Health Science)
Read more

Altmetrics

Total Views & Downloads

BROWSE