Development of MRI Scanning Technique that is Comfortable for Patients with Anxiety Disorder

Authors
Cho, Moo-SeongCho, Jae-HwanChang, Yong-MinCho, Yong-HoZeon, Seok-KilDong, Kyung-RaeChung, Woon-KwanLee, Hae-KagKim, Hyun-JuBae, Jae-YoungAhn, Jae-OukLee, Sang-Jeong
Issue Date
Dec-2011
Publisher
한국자기학회
Keywords
anxiety disorder; focusing board assembly; alpha wave; relaxation effect
Citation
Journal of Magnetics, v.16, no.4, pp 350 - 362
Pages
13
Journal Title
Journal of Magnetics
Volume
16
Number
4
Start Page
350
End Page
362
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/16076
DOI
10.4283/JMAG.2011.16.4.350
ISSN
1226-1750
2233-6656
Abstract
The principal objective of this study was to develop a scanning technique that helps patients reduce their anxiety and relax their physical tension in the MRI system. The study targeted 10 healthy persons with no medical history of anxiety disorder in the past and with no current clinically diagnosed anxiety disorder, as well as 10 patients who were currently experiencing an anxiety disorder during the MRI scanning. The focusing board assembly was self-manufactured to conduct a clinical experiment via MRI scans. As a method to confirm the efficacy of the experiment, the bio meter was used to measure brainwaves from the study targets that were divided into the normal person group (A), who felt no anxiety in the MRI system and the experimental group, (B) who did experience anxiety in the MRI system. The two groups were compared between the cases in which the focusing board assembly was used and not used after measurements were conducted using the model MRI system and the bio meter. According to the comparison and analysis results, low measurements of the alpha wave indicate highly effective relaxation of tension. In the normal person group, the alpha wave measurement showed almost no difference between cases in which the focusing board assembly was used and cases in which it was not used. In the experimental group, the alpha wave measurements were lower in cases in which the focusing board assembly was used than in cases in which the focusing board assembly was not used; this was indicative of a profound relaxation effect.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medical Sciences > Department of Medical IT Engineering > 1. Journal Articles
College of Engineering > Department of Computer Science and Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Ahn, Jae Ouk photo

Ahn, Jae Ouk
College of Software Convergence (의료IT공학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE