Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Multicenter comparison of analytical interferences of 25-OH vitamin D immunoassay and mass spectrometry methods by endogenous interferents and cross-reactivity with 3-epi-25-OH-vitamin D<sub>3</sub>open access

Authors
Lee, Joon HeeDo Seo, JongLee, KyunghoonRoh, Eun YounYun, Yeo-MinLee, Yong-WhaCho, Sung-EunSong, Junghan
Issue Date
Jan-2024
Publisher
ELSEVIER
Keywords
Vitamin D; Analytical interference; Immunoassay; Mass spectrometry; 3-epi-25-OH-Vitamin D 3
Citation
PRACTICAL LABORATORY MEDICINE, v.38
Journal Title
PRACTICAL LABORATORY MEDICINE
Volume
38
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/25927
DOI
10.1016/j.plabm.2023.e00347
ISSN
2352-5517
Abstract
Background: Vitamin D (vit-D) deficiency is highly prevalent in the Korean population, highlighting the need for accurate measurements. In this study, the interferences by endogenous and cross-reactive substances were compared between routine vit-D immunoassays and mass spectrometry (MS) methods. Methods: Two MS methods and 4 immunoassays from different manufacturers (Abbott, Beckman Coulter, Roche, Siemens) were compared. Residual samples that were icteric, lipemic, hemolyzed, high in rheumatoid factor, from myeloma patients, or patients undergoing hemodialysis were collected. Also, 4 levels of National Institute of Standards and Technology (NIST) Standard Reference Material 972a, and 12 samples serially spiked with 3-epi-25-OH-D3 were prepared. Results: Significant interferences were observed in hemolytic (Roche), icteric (Beckman and Siemens) and lipemic samples (all 4 immunoassays). Level 4 NIST material and 3-epi-25-OH-D3spiked samples induced significant cross-reactivity, yielding higher total vit-D measurements in non-epimer-separating MS methods, and both the Beckman and Roche immunoassays. Conclusion: Most observed interferences were consistent with manufacturers' claims, but overall improvement of immunoassay bias limits is required. Awareness of potential interference is important to increase the accuracy of vit-D measurements. Moreover, care is due when interpreting vit-D results of newborns, infants and less commonly, pregnant women, who are known to have physiologically high levels of the highly cross-reactive 3-epi-25-OH-D3.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Clinical Pathology > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Yong Wha photo

Lee, Yong Wha
College of Medicine (Department of Clinical Pathology)
Read more

Altmetrics

Total Views & Downloads

BROWSE