Detailed Information

Cited 0 time in webofscience Cited 7 time in scopus
Metadata Downloads

Automatic Derivation of Continuous Positive Airway Pressure Settings: Comparison with In-Laboratory Titration

Authors
Kim, HyeyunLee, MijaHwangbo, YoungYang, Kwang Ik
Issue Date
Apr-2020
Publisher
대한신경과학회
Keywords
obstructive sleep apnea; polysomnography; continuous positive airway pressure
Citation
Journal of Clinical Neurology, v.16, no.2, pp 314 - 320
Pages
7
Journal Title
Journal of Clinical Neurology
Volume
16
Number
2
Start Page
314
End Page
320
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/2964
DOI
10.3988/jcn.2020.16.2.314
ISSN
1738-6586
2005-5013
Abstract
Background and Purpose This study was designed to investigate differences in the final recommended pressure setting between that derived from an autotitrating continuous positive airway pressure (APAP) device and manual in-laboratory continuous positive airway pressure (CPAP) titration, as well as the factors that influence pressure differences in patients with obstructive sleep apnea (OSA). Methods This study enrolled 50 patients with OSA. All patients underwent both APAP titration and manual CPAP titration. We obtained the average device pressure <= 90% of the time (APAP(90)) from the downloaded manual for the APAP machine and the optimal pressure obtained by manual CPAP titration (CPAP(manual)). We placed the subjects into three groups based on the pressure difference (Delta P) obtained by subtracting CPAP(manual) from APAP(90): 1) Pr-equal (Delta P=0), 2) CPAP(manual)+ (Delta P <=-1), and 3) APAP(90)+ (Delta P >= 21). Regression analysis was conducted to identify predictive factors associated with Delta P. Results The values of APAP(90) and CPAP(manual) were 9.50 +/- 3.03 cmH(2)O and 9.48 +/- 2.71 cmH(2)O (mean SD), respectively (p=0.95). The Pr-equal, CPAP(manual)+, and APAP(90)+ groups comprised 9 (18%), 23 (46%), and 18 (36%) subjects, respectively. Regression analyses revealed that male sex [beta=3.539, 95% confidence interval (CI)=0.040-7.039], body mass index (BMI) (beta=0.186, 95% CI=0.020-0.352), and average usage per day (beta=0.768, 95% CI=0.077-1.459) were associated with Delta P. Conclusions While the mean pressure in the overall cohort did not differ significantly between APAP(90) and CPAP(manual), there was a discordance majority showing different single pressures obtained when applying the two titration methods. Being Male, having an increased BMI, and having an increased average usage per day of APAP were significantly correlated with increased Delta P in this study.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Neurology > 1. Journal Articles
College of Medicine > Department of Preventive Medicine > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yang, Kwang Ik photo

Yang, Kwang Ik
College of Medicine (Department of Neurology)
Read more

Altmetrics

Total Views & Downloads

BROWSE