Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a systematic review and meta-analysis
- Authors
- Park, Seong-Hi; Park, Young-Su
- Issue Date
- Dec-2019
- Publisher
- Lippincott Williams & Wilkins Ltd.
- Keywords
- blood pressure; blood pressure monitors; meta-analysis; sphygmomanometers; systematic review
- Citation
- Blood Pressure Monitoring, v.24, no.6, pp 265 - 276
- Pages
- 12
- Journal Title
- Blood Pressure Monitoring
- Volume
- 24
- Number
- 6
- Start Page
- 265
- End Page
- 276
- URI
- https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/3827
- DOI
- 10.1097/MBP.0000000000000412
- ISSN
- 1359-5237
1473-5725
- Abstract
- Objective: A mercury sphygmomanometer has been considered a gold standard for measuring blood pressure. However, by the Minamata Convention on Mercury, the traditional mercury sphygmomanometer is being replaced by an automated oscillometric device. This study aimed to provide scientific evidence to determine whether an automated oscillometric device can replace a mercury sphygmomanometer and if it is applicable in routine practice. Methods: MEDLINE, EMBASE, the Cochrane Library, and CINAHL were searched on 4 May 2018. Studies comparing blood pressure measurements between automated oscillometric devices and mercury sphygmomanometers were included. Study characteristics were abstracted using the evidence table, and random-effects meta-analyses were conducted. Results: Data were compiled from 24 studies comprising 47 759 subjects. The results of meta-analysis showed that automated oscillometric devices measured lower than mercury sphygmomanometers for both systolic blood pressure (mean differences -1.75 mmHg, 95% confidence intervals: -3.05 to -0.45, I-2 = 91.0%) and diastolic blood pressure (mean differences -1.20 mmHg, 95% confidence intervals: -2.16 to -0.24, I-2 = 95.0%). In sub-group analyses by manufacturer, BpTRU measured lower than the mercury sphygmomanometer and OMRON showed no difference compared to the mercury sphygmomanometer for both systolic and diastolic blood pressure, but the results differed depending on the devices. Conclusion: As a result of this review, the difference in blood pressure between the mercury sphygmomanometer and the automated oscillometric device was within 5 mmHg, but the heterogeneity between the studies was very high. The automated oscillometric devices showed differences in blood pressure results according to the manufacturer and product type.
- Files in This Item
- There are no files associated with this item.
- Appears in
Collections - College of Medicine > Department of Nursing > 1. Journal Articles
![qrcode](https://api.qrserver.com/v1/create-qr-code/?size=55x55&data=https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/3827)
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.