Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Differential Absorption Analysis of Nonmagnetic Material in the Phantom using Dual CT

Authors
Kim, Ki-YoulLee, Hae-KagCho, Jae-Hwan
Issue Date
Jun-2016
Publisher
한국자기학회
Keywords
MAR algorithm; nonmagnetic material; dual CT; CT number
Citation
Journal of Magnetics, v.21, no.2, pp 286 - 292
Pages
7
Journal Title
Journal of Magnetics
Volume
21
Number
2
Start Page
286
End Page
292
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/9078
DOI
10.4283/JMAG.2016.21.2.286
ISSN
1226-1750
2233-6656
Abstract
This study evaluates the change of computer tomography (CT) number in the case of the metal artifact reduction (MAR) algorithm, using the phantom. The images were obtained from dual CT using a gammex 467 tissue characterization phantom, which is similar to human tissues. The test method was performed by dividing pre and post MAR algorithm and measured CT values of nonmagnetic materials within the phantom. In addition, the changes of CT values for each material were compared and analyzed after measuring CT values up to 140 keV, using the spectral HU curve followed by CT scan. As a result, in the cases of N rod (trabecular bone) and E rod (trabecular bone), the CT numbers decreased as keV increasing but were constant above 90 keV. In the cases of I rod (dense bone) and K rod (dense bone), the CT numbers also decreased as keV increased but were uniform above 90 keV. The CT numbers from 40 keV to 140 keV were consistent in the cases of J rod (liver), D rod (liver), L rod (muscle), and F rod (muscle). For A rod (adipose), G rod (adipose), B rod (breast) and 0 rod (breast), the CT numbers increased as keV increased but were constant after 90 keV. The CT numbers from 40 keV to 140 keV were consistent in the case's of C rod (lung (exhale)), P rod (lung (exhale)), M rod (lung (inhale)) and H rod (lung (exhale)). Conclusively, because dual CT exhibits no changes in image quality and is able to analyze nonmagnetic materials by measuring the CT values of various materials, it will be used in the future as a useful tool for the diagnosis of lesions.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Engineering > Department of Computer Science and Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, HAE KAG photo

Lee, HAE KAG
College of Engineering (Department of Computer Science and Engineering)
Read more

Altmetrics

Total Views & Downloads

BROWSE