X-ray beam-position feedback system with easy-to-use beam-position monitor
- Authors
- Park, JY[Park, Jae Yeon]; Kim, Y[Kim, Yesul]; Lee, S[Lee, Sangsul]; Lim, J[Lim, Jun]
- Issue Date
- May-2018
- Publisher
- INT UNION CRYSTALLOGRAPHY
- Keywords
- X-ray beam-position monitor; diamond-fluorescence screen; feedback system
- Citation
- JOURNAL OF SYNCHROTRON RADIATION, v.25, no.3, pp.869 - 873
- Indexed
- SCIE
SCOPUS
- Journal Title
- JOURNAL OF SYNCHROTRON RADIATION
- Volume
- 25
- Number
- 3
- Start Page
- 869
- End Page
- 873
- URI
- https://scholarworks.bwise.kr/skku/handle/2021.sw.skku/20229
- DOI
- 10.1107/S1600577518002692
- ISSN
- 1600-5775
- Abstract
- X-ray beam-position stability is indispensable in cutting-edge experiments using synchrotron radiation. Here, for the first time, a beam-position feedback system is presented that utilizes an easy-to-use X-ray beam-position monitor incorporating a diamond-fluorescence screen. The acceptable range of the monitor is above 500 mm and the feedback system maintains the beam position within 3 mm. In addition to being inexpensive, the system has two key advantages: it works without a scale factor for position calibration, and it has no dependence on X-ray energy, X-ray intensity, beam size or beam shape.