Detailed Information

Cited 2 time in webofscience Cited 2 time in scopus
Metadata Downloads

The Incidence of Riata Defibrillator Lead Failure: a Single-Center Experienceopen access

Authors
Gwag, HB[Gwag, Hye Bin]Hwang, JK[Hwang, Jin Kyung]Park, KM[Park, Kyoung-Min]Park, SJ[Park, Seung-Jung]On, YK[On, Young Keun]Kim, JS[Kim, June Soo]
Issue Date
Oct-2017
Publisher
KOREAN ACAD MEDICAL SCIENCES
Keywords
Implantable Cardioverter-Defibrillator; Device Failure; Device Safety
Citation
JOURNAL OF KOREAN MEDICAL SCIENCE, v.32, no.10, pp.1610 - 1615
Indexed
SCIE
SCOPUS
KCI
Journal Title
JOURNAL OF KOREAN MEDICAL SCIENCE
Volume
32
Number
10
Start Page
1610
End Page
1615
URI
https://scholarworks.bwise.kr/skku/handle/2021.sw.skku/27093
DOI
10.3346/jkms.2017.32.10.1610
ISSN
1011-8934
Abstract
Riata defibrillator leads were recalled due to a high failure rate. This study measured the incidence of externalized conductor (EC) and electrical dysfunction (ED) and sought to determine the predictors of ED with Riata defibrillator leads. We enrolled patients who received Riata (R) or Riata (R) ST silicone defibrillator leads at our center between January 2003 and December 2010. The presence of EC was evaluated with chest radiography. The incidence rates were measured at < 3 years, 3-5 years, and > 5 years after lead implantation. We also investigated the rates of ED and other clinical events during the follow-up period. A total of 44 patients were analyzed. The total cumulative incidence of EC was 27.3%. During the median dwell time (80 months), the incidence of ED was 22.7%. Patients with ED were younger (46.5 vs. 56.5 years, P = 0.018) and had a higher prevalence of cardiomyopathy than those without ED (60.0% vs. 20.6%, P = 0.043). ECs were most frequently detected in patients who underwent X-ray analysis 3-5 years after lead implantation (44.4%). In contrast, ED had a bimodal incidence pattern, with peaks at 5 years (7.0%) and 9 years (5.3%). There was no difference in ED-free survival rate between patients with and without EC (P = 0.628). Given the delayed occurrence of EC and ED after implantation of Riata defibrillator leads, long-term close monitoring is critically important.
Files in This Item
There are no files associated with this item.
Appears in
Collections
Medicine > Department of Medicine > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher PARK, KYOUNG MIN photo

PARK, KYOUNG MIN
Medicine (Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE