Digital Tomosynthesis of the Chest: Comparison of Patient Exposure Dose and Image Quality between Standard Default Setting and Low Dose Setting

Authors
Hwang, HS[Hwang, Hye Sun]Chung, MJ[Chung, Myung Jin]Lee, KS[Lee, Kyung Soo]
Issue Date
May-2013
Publisher
KOREAN RADIOLOGICAL SOC
Keywords
Digital radiography; Tomographic imaging; Radiatio
Citation
KOREAN JOURNAL OF RADIOLOGY, v.14, no.3, pp.525 - 531
Indexed
SCIE
SCOPUS
KCI
Journal Title
KOREAN JOURNAL OF RADIOLOGY
Volume
14
Number
3
Start Page
525
End Page
531
URI
https://scholarworks.bwise.kr/skku/handle/2021.sw.skku/60771
DOI
10.3348/kjr.2013.14.3.525
ISSN
1229-6929
Abstract
Objective: To determine the optimum low dose (LD) digital tomosynthesis (DT) setting, and to compared the image quality of the LD DT with that of the standard default (SD) DT. Materials and Methods: Nine DT settings, by changing tube voltage, copper filter, and dose ratio, were performed for determining the LD setting. Among combinations of DT setting, a condition providing the lowest radiation dose was determined. Eighty artificial nodules Less than 1 cm in diameter (subcentimeter nodules: 40, micronodules less than 4 mm: 40) were attached to a Styrofoam and a diaphragm of the phantom. Among these, 38 nodules were located at the periphery of the lung (thin area) and 42 nodules were located at the paravertebral or sub-diaphragmatic area (thick area). Four observers counted the number of nodules detected in the thick and thin areas. The detection sensitivity in SD and LD settings were calculated separately. Data were analyzed statistically. Results: The lowest LD setting was a combination of 100 kVp, 0.3 mm additional copper filter, and a 1 : 5 dose ratio. The effective dose for the LD and SD settings were 62 mu Sv and 140 mu Sv, separately. A 56.7% dose reduction was achieved in the LD setting compared with the SD setting. Detection sensitivities were not different between the SD and the LD settings except between observers 1 and 2 for the detection of micronodules in the thick area. Conclusion: LD DT can be effective in nodule detection bigger than 4 mm without a significant decrease in image quality compared with SD DT.
Files in This Item
There are no files associated with this item.
Appears in
Collections
Medicine > Department of Medicine > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher CHUNG, MYUNG JIN photo

CHUNG, MYUNG JIN
Medicine (Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE