Measurements of beam-induced photons for particle therapy
- Authors
- Lee, KS[Lee, K. S.]; Kang, M[Kang, M.]; Jo, YM[Jo, Y. M.]; Park, SK[Park, S. K.]; Ahn, SH[Ahn, S. H.]; Han, Y[Han, Y.]
- Issue Date
- Sep-2019
- Publisher
- IOP PUBLISHING LTD
- Keywords
- Gamma camera, SPECT, PET PET/CT, coronary CT angiography (CTA); Gamma detectors (scintillators, CZT, HPG, HgI etc); Inspection with gamma rays; Instrumentation for hadron therapy
- Citation
- JOURNAL OF INSTRUMENTATION, v.14, no.9
- Indexed
- SCIE
SCOPUS
- Journal Title
- JOURNAL OF INSTRUMENTATION
- Volume
- 14
- Number
- 9
- URI
- https://scholarworks.bwise.kr/skku/handle/2021.sw.skku/8887
- DOI
- 10.1088/1748-0221/14/09/C09034
- ISSN
- 1748-0221
- Abstract
- Resistive plate chambers are studied for in-vivo beam verification in particle therapy. A four-gap glass RPC is constructed with 1.1-mm-thick floating glass and tested with 662-keV gamma rays emitted from a 5-GBq Cs-137 source, as well as high-energy photons induced by 140-MeV raster-scan-mode proton beams provided by the Samsung Proton Therapy Center. The desired 3 mm resolution of the RPC detector is confirmed by reconstructing gamma transmission images for the cesium gammas gathered from the data. Using the described RPC detector and a 10-cm-thick lead collimator, we obtain two-dimensional gamma-ray images for 140-MeV proton beams and confirm the detector technology proposed in the present research.
- Files in This Item
- There are no files associated with this item.
- Appears in
Collections - Medicine > Department of Medicine > 1. Journal Articles
![qrcode](https://api.qrserver.com/v1/create-qr-code/?size=55x55&data=https://scholarworks.bwise.kr/skku/handle/2021.sw.skku/8887)
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.