Flexible and Implantable Capacitive Microelectrode for Bio-potential Acquisition

Authors
Lee, Seung MinByeon, Hang JinKim, Bong HoonLee, JungynpJeong, Ji YoonLee, Joong HoonMoon, Jin-HeePark, CheolsooChoi, HyukLee, Sang-HoonLee, Kwang-Ho
Issue Date
Jun-2017
Publisher
KOREAN BIOCHIP SOCIETY-KBCS
Keywords
Electrode: Polydimethylsiloxane (PDMS); Gold/Titanium (Au/Ti); Electrocardiography (ECG); Capacitive Coupling; Bio-potential
Citation
BIOCHIP JOURNAL, v.11, no.2, pp.153 - 163
Journal Title
BIOCHIP JOURNAL
Volume
11
Number
2
Start Page
153
End Page
163
URI
http://scholarworks.bwise.kr/ssu/handle/2018.sw.ssu/39301
DOI
10.1007/s13206-017-1304-y
ISSN
1976-0280
Abstract
Electrodes should be adhered onto the body steadily under motion, and implanted stably into the tissue without any damages while maintaining conformal contact. Although most electrodes are fabricated with biocompatible materials, they should be shielded from tissues to prevent mechanical delamination from the device itself and to avoid adverse effects such as irritation, allergic reactions, or inflammation. Herein, we demonstrate a simple process for the development of a flexible and implantable capacitive electrode based on a minimal incision accessible design with polyimide (PI) and Gold/Titanium (Au/Ti) layers and completely encapsulated in a polydimethylsiloxane (PDMS) substrate. Electrodes of three different sizes (recording site diameters of 1.8 mm, 2.8 mm, and 3.8 mm, respectively) were fabricated and examined in this work. Electrocardiography (ECG) was recorded in the dorsal area of the rat for 4 weeks for biological signal checkup. We obtained stable and robust ECG signals owing-to the intrinsic property of capacitive coupling, with almost no leakage current compared to the direct contact electrode for the applied current over the range of 0 to 10 mA. These results indicate that our electrode can be used to detect bio-signals effectively in the long term, and can play a role in electroceuticals in the near future.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Engineering > ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE