유량측정을 위한 초음파 전파경로 비교Comparison of Ultrasonic Paths for Flow Rate Measurement

Other Titles
Comparison of Ultrasonic Paths for Flow Rate Measurement
Authors
김주완박춘광김진오박두식
Issue Date
Jul-2015
Publisher
한국소음진동공학회
Keywords
Ultrasound; Path; Flow Rate; Transducer; Transit Time; 초음파; 경로; 유량; 변환기; 전파시간
Citation
한국소음진동공학회논문집, v.25, no.7, pp.455 - 461
Journal Title
한국소음진동공학회논문집
Volume
25
Number
7
Start Page
455
End Page
461
URI
http://scholarworks.bwise.kr/ssu/handle/2018.sw.ssu/9335
DOI
10.5050/KSNVE.2015.25.7.455
ISSN
1598-2785
Abstract
The paper deals with the accuracy comparison between two kinds of ultrasonic paths for flow rate measurement. In the Z-path, two transducers are installed on the opposite sides of each other on a pipe, and the ultrasonic waves generated at one transducer propagate to arrive at the other one only by refraction. In the V-path, two transducers are installed on one side of a pipe, and the ultrasonic waves reflected at the inner wall of the pipe are received. Transit times were confirmed to identify the propagation paths by comparing the theoretically calculated results and measured ones. The flow rate measurements with two kinds of ultrasonic paths appeared very similar. It would be possible to select either of the paths by considering the advantages and disadvantages.
Files in This Item
Go to Link
Appears in
Collections
College of Engineering > ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Jin Oh photo

Kim, Jin Oh
College of Engineering (School of Mechanical Engineering)
Read more

Altmetrics

Total Views & Downloads

BROWSE