Detailed Information

Cited 17 time in webofscience Cited 19 time in scopus
Metadata Downloads

Less perilaryngeal gas leakage with SLIPA(TM) than with LMA-ProSeal(TM) in paralyzed patients

Authors
Woo, Young CheolCha, Su ManKang, HyunBaek, Chong WhaJung, Yong HunKim, Jin YunKoo, Gill HoiPark, Sun GyooKim, Seong Deok
Issue Date
Jan-2011
Publisher
SPRINGER
Citation
CANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE, v.58, no.1, pp 48 - 54
Pages
7
Journal Title
CANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE
Volume
58
Number
1
Start Page
48
End Page
54
URI
https://scholarworks.bwise.kr/cau/handle/2019.sw.cau/21840
DOI
10.1007/s12630-010-9412-3
ISSN
0832-610X
1496-8975
Abstract
The aim of this study was to compare the Streamlined Liner of the Pharynx Airway (SLIPA (TM)) with the ProSeal Laryngeal Mask Airway (LMA-ProSeal (TM)) in mechanically ventilated paralyzed patients undergoing laparoscopic gynecologic surgery. One hundred and one patients were allocated randomly to SLIPA (n = 50) or to LMA-ProSeal (n = 51) treatment groups. After induction of general anesthesia and insertion of the assigned supralaryngeal airway (SLA) device, we made note of the occurrence of any gastric insufflation and perilaryngeal leakage. We then evaluated the anatomical fit of the SLA device using a fibreoptic bronchoscope, and we assessed the airway sealing pressure and respiratory mechanics with change in head position and during peritoneal insufflation. After surgery, we evaluated the severity of postoperative sore throat and the presence of blood or regurgitated fluid on the SLA device. The insertion success rate, gastric insufflation, perilaryngeal leakage, anatomical fit, airway sealing pressure, respiratory mechanics, severity of sore throat, and incidence of blood and regurgitated fluid on the device were similar between the two groups. The incidence of perilaryngeal leakage with changes in the patient's head position was lower with the SLIPA group than with the LMA-ProSeal group (3/50 vs 11/51, respectively; P = 0.026). During peritoneal insufflation, perilaryngeal leakage did not occur with the SLIPA but occurred in four cases with the LMA-ProSeal (P = 0.045). Both the SLIPA and the LMA-ProSeal can be used effectively and without severe complications in paralyzed patients undergoing laparoscopic gynecological surgery. However, the SLIPA offers the advantage of less perilaryngeal gas leakage than the LMA-ProSeal with change in head position and during insufflation of the peritoneal cavity. This trial is registered with ANZCTR (ACTRN12609000914268).
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > College of Medicine > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kang, Hyun photo

Kang, Hyun
의과대학 (의학부(임상-서울))
Read more

Altmetrics

Total Views & Downloads

BROWSE