Effects of bacteria and yeast on WBC counting in three automated hematology counters

Authors
Kim, Hye RyounPark, Bo Rae G.Lee, Mi Kyung
Issue Date
Jul-2008
Publisher
SPRINGER
Keywords
ADVIA 120; 2120; Sysmex XE-2100; Coulter LH 750; candidemia
Citation
ANNALS OF HEMATOLOGY, v.87, no.7, pp 557 - 562
Pages
6
Journal Title
ANNALS OF HEMATOLOGY
Volume
87
Number
7
Start Page
557
End Page
562
URI
https://scholarworks.bwise.kr/cau/handle/2019.sw.cau/23712
DOI
10.1007/s00277-008-0464-1
ISSN
0939-5555
1432-0584
Abstract
Bacteria or yeast may be observed on peripheral blood smears and may lead to spuriously elevated platelet counts. They have been reported to disturb the white blood cell (WBC) differential count if they clumped together, and a large number of such microorganisms have been shown to increase WBC counts. The purpose of this study was to evaluate the spurious rise in WBC counts according to species of microorganisms and automated hematology analyzers. The species we selected were Staphylococcus aureus, Escherichia coli, Candida albicans, C. tropicalis, C. krusei, C. dubliniensis, C. glabrata, and C. parapsilosis. We investigated the effects of bacteria and yeast on peripheral blood samples by the ADVIA 120/2120 Hematology System, Sysmex XE-2100 (TOA Medical Electronics, Kobe, Japan) and Coulter LH 750 (Beckman Coulter, Miami, FL, USA). C. albicans, C. tropicalis, C. krusei, and C. dubliniensis had an overt effect on the WBC count at concentrations of up to 1-5 x 10(7) colony-forming units (CFU)/mL in three automated cell counters, and C. glabrata and C. parapsilosis, when present at concentrations of 1-5 x 10(8) CFU/mL, caused a significant increase in the WBC count obtained by the Sysmex XE-2100 but not by the ADVIA 120/2120 system and Coulter LH 750 (p < 0.05). In conclusion, yeast may influence the results of peripheral blood smears only when the yeast concentration is unusually high. The results differed among analyzers and among species of yeast. Hematologists should be aware that samples containing bacteria and yeast may give erroneously high WBC counts and differential leukocyte counts and should review the peripheral blood smear by microscopy.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > College of Medicine > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Mi-Kyung photo

Lee, Mi-Kyung
의과대학 (의학부(임상-서울))
Read more

Altmetrics

Total Views & Downloads

BROWSE