Characteristics of a transverse-type Compton-suppressed gamma-ray detector
- Authors
- Lee, CS; Kwon, YK; Lee, JH; Kim, JH; Kim, JY; Moon, JY
- Issue Date
- May-2002
- Publisher
- KOREAN PHYSICAL SOC
- Keywords
- gamma-ray detector; Compton-suppressed detector; BGO; cyclotron
- Citation
- JOURNAL OF THE KOREAN PHYSICAL SOCIETY, v.40, no.5, pp 807 - 811
- Pages
- 5
- Journal Title
- JOURNAL OF THE KOREAN PHYSICAL SOCIETY
- Volume
- 40
- Number
- 5
- Start Page
- 807
- End Page
- 811
- URI
- https://scholarworks.bwise.kr/cau/handle/2019.sw.cau/25119
- ISSN
- 0374-4884
- Abstract
- The performance of a Compton-suppressed gamma-ray detector using a transverse escape-suppression shield of BGO was analyzed by measuring two relevant factors, the suppression ratio and the peak-to-total ratio. The suppression ratio varied from 2 to 6, depending on gamma-ray energy as two standard sources. Co-60 and Cs-137, were used. The peak-to-total ratio was improved by a factor of two or more when the escape shield was used. The suppression ratio of our detector was shown to be comparable to that of a Stony-Brook detector of a similar type.