Repeated endotracheal tube cuff tears during nasotracheal intubation due to nasal cavity orthodontic micro-implant - A case report -
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Seo Su Hyun | - |
dc.contributor.author | Lee Jun Mo | - |
dc.contributor.author | Lee Je Jin | - |
dc.contributor.author | Ahn Eun Jin | - |
dc.contributor.author | Choi Geun Joo | - |
dc.contributor.author | Kang Hyun | - |
dc.date.accessioned | 2021-11-17T01:40:18Z | - |
dc.date.available | 2021-11-17T01:40:18Z | - |
dc.date.issued | 2021-10 | - |
dc.identifier.issn | 1975-5171 | - |
dc.identifier.issn | 2383-7977 | - |
dc.identifier.uri | https://scholarworks.bwise.kr/cau/handle/2019.sw.cau/51492 | - |
dc.description.abstract | Background: Nasotracheal intubation is generally performed for intraoral surgery.Case: A 34-year-old female patient who underwent orthognathic surgery exhibited repeated endotracheal tube cuff tears during nasotracheal intubation. After intubation, leaks developed, and torn endotracheal cuff was observed in the removed endotracheal tube. Subsequently, re-intubation through the same nasal cavity was performed immediately, but leakage from the torn endotracheal tube cuff was re-observed. A leakage test of the extubated tube revealed air bubbles and leaks near the tube cuff due to the tear. Nasotracheal intubation was performed through the other nasal cavity, and there were no leakage findings or abnormalities. During the course of the surgery, the surgeon noticed that the orthodontic micro-implant deposited in the mid-tube cavity was exposed to the nasal cavity. Conclusions: We aimed to emphasize caution and discuss the possibility that orthodontic micro-implants that are not confirmed during preoperative evaluation may cause repeated endotracheal tube cuff tears. | - |
dc.format.extent | 5 | - |
dc.language | 영어 | - |
dc.language.iso | ENG | - |
dc.publisher | 대한마취통증의학회 | - |
dc.title | Repeated endotracheal tube cuff tears during nasotracheal intubation due to nasal cavity orthodontic micro-implant - A case report - | - |
dc.type | Article | - |
dc.identifier.doi | 10.17085/apm.21015 | - |
dc.identifier.bibliographicCitation | Anesthesia and Pain Medicine, v.16, no.4, pp 398 - 402 | - |
dc.identifier.kciid | ART002768604 | - |
dc.description.isOpenAccess | Y | - |
dc.identifier.scopusid | 2-s2.0-85167922699 | - |
dc.citation.endPage | 402 | - |
dc.citation.number | 4 | - |
dc.citation.startPage | 398 | - |
dc.citation.title | Anesthesia and Pain Medicine | - |
dc.citation.volume | 16 | - |
dc.publisher.location | 대한민국 | - |
dc.subject.keywordAuthor | Cuff leakage | - |
dc.subject.keywordAuthor | Dental implants | - |
dc.subject.keywordAuthor | Intratracheal intubation | - |
dc.subject.keywordAuthor | Nasotracheal intubation | - |
dc.subject.keywordAuthor | Orthognathic surgery | - |
dc.description.journalRegisteredClass | kci | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
84, Heukseok-ro, Dongjak-gu, Seoul, Republic of Korea (06974)02-820-6194
COPYRIGHT 2019 Chung-Ang University All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.