Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

External Validation of Deep Learning Algorithm for Detecting and Visualizing Femoral Neck Fracture Including Displaced and Non-displaced Fracture on Plain X-ray

Authors
Bae, JunwonYu, SangjoonOh, JaehoonKim, Tae HyunChung, Jae HoByun, HayoungYoon, Myeong SeongAhn, ChiwonLee, Dong Keon
Issue Date
Oct-2021
Publisher
SPRINGER
Keywords
Femur; Fracture; Deep learning; Machine learning; Artificial intelligence; AI
Citation
JOURNAL OF DIGITAL IMAGING, v.34, no.5, pp 1099 - 1109
Pages
11
Journal Title
JOURNAL OF DIGITAL IMAGING
Volume
34
Number
5
Start Page
1099
End Page
1109
URI
https://scholarworks.bwise.kr/cau/handle/2019.sw.cau/62167
DOI
10.1007/s10278-021-00499-2
ISSN
0897-1889
1618-727X
Abstract
This study aimed to develop a method for detection of femoral neck fracture (FNF) including displaced and non-displaced fractures using convolutional neural network (CNN) with plain X-ray and to validate its use across hospitals through internal and external validation sets. This is a retrospective study using hip and pelvic anteroposterior films for training and detecting femoral neck fracture through residual neural network (ResNet) 18 with convolutional block attention module (CBAM) + +. The study was performed at two tertiary hospitals between February and May 2020 and used data from January 2005 to December 2018. Our primary outcome was favorable performance for diagnosis of femoral neck fracture from negative studies in our dataset. We described the outcomes as area under the receiver operating characteristic curve (AUC), accuracy, Youden index, sensitivity, and specificity. A total of 4,189 images that contained 1,109 positive images (332 non-displaced and 777 displaced) and 3,080 negative images were collected from two hospitals. The test values after training with one hospital dataset were 0.999 AUC, 0.986 accuracy, 0.960 Youden index, and 0.966 sensitivity, and 0.993 specificity. Values of external validation with the other hospital dataset were 0.977, 0.971, 0.920, 0.939, and 0.982, respectively. Values of merged hospital datasets were 0.987, 0.983, 0.960, 0.973, and 0.987, respectively. A CNN algorithm for FNF detection in both displaced and non-displaced fractures using plain X-rays could be used in other hospitals to screen for FNF after training with images from the hospital of interest.
Files in This Item
There are no files associated with this item.
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Ahn, Chiwon photo

Ahn, Chiwon
의과대학 (의학부(임상-서울))
Read more

Altmetrics

Total Views & Downloads

BROWSE