Detailed Information

Cited 3 time in webofscience Cited 2 time in scopus
Metadata Downloads

Pre-warming the Streamlined Liner of the Pharynx Airway (SLIPA (TM)) improves fitting to the laryngeal structure: a randomized, double-blind studyopen access

Authors
Kang, HyunKim, Dong RimJung, Yong HunBaek, Chong WhaPark, Yong HeeOh, Jong InKim, Won JoongChoi, Geun Joo
Issue Date
Nov-2015
Publisher
BIOMED CENTRAL LTD
Keywords
Pre-warming; SLIPA; Oropharyngeal leak pressure; Supraglottic airway devices
Citation
BMC ANESTHESIOLOGY, v.15, no.1
Journal Title
BMC ANESTHESIOLOGY
Volume
15
Number
1
URI
https://scholarworks.bwise.kr/cau/handle/2019.sw.cau/8905
DOI
10.1186/s12871-015-0151-4
ISSN
1471-2253
Abstract
Background: The Streamlined Liner of the Pharynx Airway (SLIPA), a type of supraglottic airway, has a non-inflatable cuff that softens at body temperature to fit the laryngeal structure. We investigated whether pre-warming of SLIPA to body temperature may improve insertion parameters. Methods: Ninety adult patients were assigned equally randomized to either Group W or Group R. Anesthesia was induced using propofol, fentanyl, and rocuronium. In Group W, the SLIPA was warmed to 37 degrees C before insertion, whereas in Group R, it was inserted at room temperature. The insertion time, oropharyngeal leak pressure, postoperative throat pain, blood staining, regurgitation, number of attempts at insertion, and difficulty of insertion were compared between the two groups. Results: The insertion time was shorter in Group W than in Group R (3.60 [3.15-4.06] s vs. 6.00 [4.45-7.50] s; P < 0.001). Oropharyngeal leak pressure from the time of insertion until 3 min after insertion was significantly higher in Group W than in Group R (P < 0.05). Postoperative throat pain, measured using the visual analog scale, was lower in Group W than in Group R (0.00 [0.00 2.50] vs. 2.00 [0.00 4.50]; P = 0.006). The difficulty of insertion was lower in Group W than in Group R (P < 0.004). There were no significant differences in terms of blood staining, regurgitation, and number of attempts. Conclusions: Pre-warming the SLIPA to body temperature has significant benefits compared to maintaining the device at room temperature. Specifically, insertion was easier, both insertion and fitting to the laryngeal structure could be performed more quickly, and the incidence of sore throat was reduced.
Files in This Item
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Jung, Yong Hun photo

Jung, Yong Hun
의과대학 (의학부(임상-광명))
Read more

Altmetrics

Total Views & Downloads

BROWSE