4 MHz Electrosurgical Generator System for Wide Load Impedance Range with SiC-Based Full Bridge Inverter
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Yoon, Byeongcheol | - |
dc.contributor.author | Hwang, Sungwoon | - |
dc.contributor.author | Lee, Seokmin | - |
dc.contributor.author | Jeong, Byongchang | - |
dc.contributor.author | Kim, Jongchan | - |
dc.contributor.author | Kim, Sungmin | - |
dc.contributor.author | Kim, Junghyun | - |
dc.date.accessioned | 2023-07-05T05:30:55Z | - |
dc.date.available | 2023-07-05T05:30:55Z | - |
dc.date.issued | 2024-01 | - |
dc.identifier.issn | 0278-0046 | - |
dc.identifier.issn | 1557-9948 | - |
dc.identifier.uri | https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/112847 | - |
dc.description.abstract | A power conversion system to drive a high-frequency electrosurgical generator (ESG) is proposed. The proposed system consists of a 4 MHz full-bridge (FB) inverter capable of ensuring power in a wide load impedance range and implementing energy control functions in accordance with the ESG characteristics. Unlike conventional inverter systems, ESG cannot deliver energy directly to the load, but through an instrument, which makes it difficult to ensure power. Thus, impedance matching is essential to ensure power in ESG. For ensuring power, impedance matching is realized using a resonance filter while considering instrument induced parasitic components that cannot be ignored at high frequencies. The inverter is composed of silicon carbide (SiC) suitable for the MHz switching operation. A prototype is to validate the proposed system. Experimental test of the prototype that it can provide a load power of more than 100 W for a tissue model with impedance range of 5-to-800 Ω. | - |
dc.format.extent | 9 | - |
dc.language | 영어 | - |
dc.language.iso | ENG | - |
dc.publisher | Institute of Electrical and Electronics Engineers | - |
dc.title | 4 MHz Electrosurgical Generator System for Wide Load Impedance Range with SiC-Based Full Bridge Inverter | - |
dc.type | Article | - |
dc.publisher.location | 미국 | - |
dc.identifier.doi | 10.1109/TIE.2023.3247795 | - |
dc.identifier.scopusid | 2-s2.0-85149387503 | - |
dc.identifier.wosid | 001037241000031 | - |
dc.identifier.bibliographicCitation | IEEE Transactions on Industrial Electronics, v.71, no.1, pp 1 - 9 | - |
dc.citation.title | IEEE Transactions on Industrial Electronics | - |
dc.citation.volume | 71 | - |
dc.citation.number | 1 | - |
dc.citation.startPage | 1 | - |
dc.citation.endPage | 9 | - |
dc.type.docType | Article | - |
dc.description.isOpenAccess | N | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.relation.journalResearchArea | Automation & Control Systems | - |
dc.relation.journalResearchArea | Engineering | - |
dc.relation.journalResearchArea | Instruments & Instrumentation | - |
dc.relation.journalWebOfScienceCategory | Automation & Control Systems | - |
dc.relation.journalWebOfScienceCategory | Engineering, Electrical & Electronic | - |
dc.relation.journalWebOfScienceCategory | Instruments & Instrumentation | - |
dc.subject.keywordAuthor | High-frequency electrosurgical generator (ESG) | - |
dc.subject.keywordAuthor | high-frequency full-bridge (FB) inverter | - |
dc.subject.keywordAuthor | impedance matching | - |
dc.subject.keywordAuthor | lookup table (LUT) | - |
dc.subject.keywordAuthor | resonance filter | - |
dc.subject.keywordAuthor | variable impedance | - |
dc.subject.keywordAuthor | wide load impedance range | - |
dc.identifier.url | https://ieeexplore.ieee.org/document/10054584 | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
55 Hanyangdeahak-ro, Sangnok-gu, Ansan, Gyeonggi-do, 15588, Korea+82-31-400-4269 sweetbrain@hanyang.ac.kr
COPYRIGHT © 2021 HANYANG UNIVERSITY. ALL RIGHTS RESERVED.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.