Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

3D-Printed Patient-Specific Circles of Willis with an Intracranial Aneurysm and their Application to Neurointerventional Endovascular Simulation

Authors
Kim, YeonwooYang, HyeondongCho, Kwang-ChunKim, Jung-JaeKim, Yong BaeOh, Je Hoon
Issue Date
Aug-2023
Publisher
JOHN WILEY & SONS INC
Keywords
cerebrovascular replica; circle of Willis; intracranial aneurysms; layer-by-layer coatings; sneurointerventional endovascular simulator
Citation
Advanced Materials Technologies, v.8, no.15, pp 1 - 13
Pages
13
Indexed
SCIE
SCOPUS
Journal Title
Advanced Materials Technologies
Volume
8
Number
15
Start Page
1
End Page
13
URI
https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/112996
DOI
10.1002/admt.202201783
ISSN
2365-709X
Abstract
Cerebrovascular replicas are effective platforms that can simulate endovascular interventions in the treatment of intracranial aneurysm (IA), and an IA typically forms in the tortuously structured cerebral artery circulation known as the circle of Willis (CoW). Effective and reliable endovascular procedure simulation before actual clinical treatment is greatly helpful for neurosurgeons, but the simulation performance depends on how identical a CoW replica is to a patient's CoW. Herein, a tubular, transparent, and patient-specific CoW replica is fabricated with realistic dimensions and elasticity using a combination of 3D printing and dip coating technique. First, a real-scale CoW core substrate is constructed using an extrusion-based 3D printing technology, then the thickness of the vascular wall is manipulated by coating the core substrate layer by layer, and finally the core substrate is dissolved to produce a tubular CoW structure. A liquid-assisted dip coating method is utilized and optimized for uniform and high-quality coating layers on the tortuous CoW core substrate. Moreover, the effectiveness of the CoW replica is demonstrated through an in-house neurointerventional endovascular simulator. The proposed method paves a new way for practical and reliable endovascular simulations, which would help significantly improve the clinical outcomes of current IA treatments.
Files in This Item
Go to Link
Appears in
Collections
COLLEGE OF ENGINEERING SCIENCES > DEPARTMENT OF MECHANICAL ENGINEERING > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Oh, Je Hoon photo

Oh, Je Hoon
ERICA 공학대학 (DEPARTMENT OF MECHANICAL ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE