Continuous electrochemical detection of hydrogen peroxide by Au-Ag bimetallic nanoparticles in microfluidic devices

Authors
Ko, EunaVan-Khue TranGeng, YanfangChung, Woo SungPark, Chan HoKim, Min KiJin, Ga HyunSeong, Gi Hun
Issue Date
May-2017
Publisher
ELSEVIER SCIENCE SA
Keywords
Au-Ag nanoparticles; Microfluidic device; Electrocatalysis; Hydrogen peroxide; Single-walled carbon nanotube
Citation
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, v.792, pp 72 - 78
Pages
7
Indexed
SCI
SCIE
SCOPUS
Journal Title
JOURNAL OF ELECTROANALYTICAL CHEMISTRY
Volume
792
Start Page
72
End Page
78
URI
https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/9609
DOI
10.1016/j.jelechem.2017.03.027
ISSN
1572-6657
1873-2569
Abstract
We developed electrochemical microfluidic devices containing Au-Ag bimetallic nanoparticles on single-walled carbon nanotube (SWCNT) electrodes for precise, rapid and non-enzymatic electrochemical detection of hydrogen peroxide (H2O2). A homogeneous SWCNT layer on a glass substrate was patterned by a standard photolithography method and an O-2-plasma technique. The patterned SWCNT layers were used as working electrode, counter electrode, and reference electrode in a 1 mm wide and 23 pm deep microchannel. To enhance the electrochemical catalytic properties of the working electrode, Au covered Ag nanoparticles (NPs) were electrochemically deposited on the SWCNT working electrode with current streaming solution into a microfluidic channel. The.AgNPs had catalytic activity to improve the reduction of H2O2 combined with AuNPs, which could support catalytic reaction of AgNPs. The reduction in H2O2 in fabricated devices was monitored at an optimized condition and was linear with H2O2 concentration, resulting in a sensitivity of 13.1 mu A.cm(-2).mM(-1), a correlation coefficient of r(2) = 0.985, and limit of detection 26.8 mu M (S/N = 3). For the analysis of real samples, diluted antiseptic solutions containing 3 wt% H2O2 were used to determine the concentration of H2O2 in the pharmaceutical sample. The recovery values ranged from 97% to 107%, and relative standard deviations were below 5%. We expect that the developed microfluidic sensor is a prospective electrochemical platform for accurate, reproducible non-enzymatic detection of H2O2. (c) 2017 Elsevier B.V. All rights reserved.
Files in This Item
Go to Link
Appears in
Collections
COLLEGE OF ENGINEERING SCIENCES > DEPARTMENT OF BIONANO ENGINEERING > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Seong, Gi Hun photo

Seong, Gi Hun
ERICA 공학대학 (DEPARTMENT OF BIONANO ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE