Measurement of Fractional Exhaled Nitric Oxide in Adults: Comparison of Two Different Analyzers (NIOX VERO and NObreath)

Authors
강성윤이상민이상표
Issue Date
Jul-2021
Publisher
대한결핵및호흡기학회
Keywords
Exhalation; Nitric Oxide; Biomarkers; Respiratory Disease
Citation
Tuberculosis and Respiratory Diseases, v.84, no.3, pp.182 - 187
Journal Title
Tuberculosis and Respiratory Diseases
Volume
84
Number
3
Start Page
182
End Page
187
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/81691
DOI
10.4046/trd.2020.0137
ISSN
1738-3536
Abstract
Background: Fractional exhaled nitric oxide (FeNO) is a non-invasive marker for eosinophilic airway inflammation and a good predictor of response to corticosteroids. There is a need for a reliable and accurate measurement method, as FeNO measurements have been widely used in clinical practice. Our study aimed to compare two FeNO analyzers and derive a conversion equation for FeNO measurements in adults. Methods: We included 99 participants who had chief complaints of chronic cough and difficulty in breathing. The participants underwent concurrent FeNO measurement using NIOX VERO (Circassia AB) and NObreath (Bedfont). We compared the values of the two devices and analyzed their correlation and agreement. We then formulated an equation to convert FeNO values measured by NObreath into those obtained by NIOX VERO. Results: The mean age of the participants was 51.2±17.1 years, with a female predominance (58.6%). Approximately 60% of the participants had asthma. The FeNO level measured by NIOX VERO (median, 27; interquartile range [IQR], 15–45) was significantly lower than that measured by NObreath (median, 38; IQR, 22–58; p<0.001). There was a strong positive correlation between the two devices (r=0.779, p<0.001). Additionally, Bland-Altman plots and intraclass correlation coefficient demonstrated a good agreement. Using linear regression, we derived the following conversion equation: natural log (Ln) (NObreath)=0.728×Ln (NIOX VERO)+1.244. Conclusion: The FeNO values of NIOX VERO and NObreath were in good agreement and had positive correlations. Our proposed conversion equation could help assess the accuracy of the two analyzers.
Files in This Item
There are no files associated with this item.
Appears in
Collections
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kang, Sung Yoon photo

Kang, Sung Yoon
College of Medicine (Department of Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE