Validation of intra- and interobserver reproducibility of shearwave elastography: Phantom study

Authors
Mun, Han SongChoi, Seon HyeongKook, Shin HoChoi, YoonjungJeong, Woo KyoungKim, Yongsoo
Issue Date
Jul-2013
Publisher
ELSEVIER
Keywords
Ultrasound; Elastography; Shearwave elastography; Reproducibility; Phantom
Citation
ULTRASONICS, v.53, no.5, pp.1039 - 1043
Indexed
SCIE
SCOPUS
Journal Title
ULTRASONICS
Volume
53
Number
5
Start Page
1039
End Page
1043
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/162450
DOI
10.1016/j.ultras.2013.01.013
ISSN
0041-624X
Abstract
Purpose: To assess intra- and interobserver reproducibility of quantitative ultrasound (US) elastography for breast masses using shearwave elastography (SWE). Materials and methods: The US elastography phantom was used. This training phantom contained 12 masses with dimensions from 6 to 11 mm and varying B-mode (hypo- iso- and hyperechoic) characteristics and viscoelastic properties varying from about 50 to 250 kPa. Four attending radiologists participated as operators and used Supersonic Imaging 15-4 MHz array to acquire images. In each session, B-mode US features were assessed once and US elastography values were assessed twice in a consecutive and distinct manner. A total of three identical sessions of data acquisition were performed at 2-week intervals. For categorical variables such as BI-RADS features of mass on B-mode image and qualitative SWE features, kappa (k) values were calculated by the generalized linear mixed model. For continuous variables, kPa, and ratio, intraclass correlation coefficients (ICCs) were calculated. Results: Interobserver agreement for SWE image similarity was moderate (k = 0.57). With respect to quantitative SWE values, intraobserver ICCs were 0.65, 0.77, 0.92, and 0.91 for maximum elasticity; 0.70, 0.83, 0.94, and 0.94 for mean elasticity; and 0.67, 0.83, 0.92, and 0.92 for elasticity ratio for operators 1, 2, 3, and 4, respectively. Interobserver reproducibility showed good agreement with ICC values of 0.77 for maximum elasticity, 0.82 for mean elasticity, and 0.79 for elasticity ratio. Conclusion: Quantitative US elastography for breast masses was highly reproducible with good agreement across operators. With respect to intraobserver reproducibility, ICC values varied but three out of four operators showed good agreement. (C) 2013 Elsevier B. V. All rights reserved.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 영상의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Yong Soo photo

Kim, Yong Soo
COLLEGE OF MEDICINE (DEPARTMENT OF RADIOLOGY)
Read more

Altmetrics

Total Views & Downloads

BROWSE