Determination of the beam quality correction factor for the microLion chamber in a clinical photon beam
- Authors
- Choi, Sang Hyoun; Kim, Chan Hyeong; Do Huh, Hyun; Kim, Kum-Bae; Kim, SeongHoon
- Issue Date
- Jan-2013
- Publisher
- 한국물리학회
- Keywords
- Beam quality correction factor; MicroLion chamber; Monte Carlo simulation
- Citation
- Journal of the Korean Physical Society, v.62, no.1, pp 152 - 158
- Pages
- 7
- Indexed
- SCI
SCIE
SCOPUS
KCI
- Journal Title
- Journal of the Korean Physical Society
- Volume
- 62
- Number
- 1
- Start Page
- 152
- End Page
- 158
- URI
- https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/163678
- DOI
- 10.3938/jkps.62.152
- ISSN
- 0374-4884
1976-8524
- Abstract
- The mcroLion chamber, which has a small volume (0.0017cc) and a high sensitivity suitable for small field dosimetry, is an ionization chamber with liquid (isooctane) filling the cavity volume. In the present work, we used two alternative ways to estimate the beam quality correction factor of the microLion chamber in the 6 MV photon beam from a Clinac iX accelerator (Varian, USA); Monte Carlo simulations and experimental methods. The Monte Carlo calculation (1.0183 +/- 0.58%) agreed to within 0.56% with the experimental determination (1.024 +/- 0.58%). The results from the two methods were in good agreement within acceptable uncertainties. We think that the averaged factor, 1.0212, could be used for the microLion chamber in reference dosimetry.
- Files in This Item
-
Go to Link
- Appears in
Collections - 서울 공과대학 > 서울 원자력공학과 > 1. Journal Articles

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.